Recall Z-0537-2022

Dixi Medical USA · initiated December 9, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Codes / lots: All lots UDI: 03664539000077

Reason for recall

Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

FDA-determined cause: Under Investigation by firm

Action taken

Dixi Medical issued Urgent Medical Device Letter on 12/9/21. Letter states reason for recall, health risk and action to take: 1. Examine your inventory for the affected product. Any available stock must be quarantined immediately and placed in a secure location to prevent further usage. 2. Please use the response form to record and return the stock back to the manufacturer. 3. The returned stock will be replaced with compliant products. 4. Please complete and return the attached Customer Response Form to Ann Cirner, ann.cirner@diximedus.com (248)845-3494, even if you do not have affected product on hand. 5. Please forward this information to all your staff that may use the MICRODEEP during a SEEG procedure in your healthcare facility. 6. If you have further distributed the product, please forward this notification to any downstream accounts. The removal and return of affected products will be carried out by the hospital staff. Your assistance is appreciated and necessary to prevent a potential patient harm. If you have any questions, call Ann Cirner, ann.cirner@diximedus.com (248)845-3494

Details

Recall number
Z-0537-2022
Initiated
December 9, 2021
Terminated
September 8, 2023
Product code
GZL – Electrode, Depth
Quantity
2720 units
Distribution
US Nationwide distribution.
Recalling firm
Dixi Medical USA, Plymouth, MI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0534-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0535-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0536-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0538-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0539-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0540-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0541-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated
Z-0542-2022 Class IIThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a…Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the… 12/09/2021Terminated