Recall Z-0544-2020

Abbott Gmbh & Co. KG · initiated September 16, 2019

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Accessory kit, Alinity i containing Alinity ci Level Sensor, Bulk Solution

Codes / lots: List Number 03R66 03, All Serial Numbers

Reason for recall

Potential reliability issue with the Alinity ci series Level Sensor, Bulk Solution. The development of environmental stress on the bulk solution level sensor can cause cracks.

FDA-determined cause: Device Design

Action taken

Abbott GmbH & Co. KG notified customers on about 09/16/2019, via "Product Correction Urgent Immediate Action Required" letter. Customers were informed that the development of environmental stress on the bulk solution level sensor can cause cracks. Cracks can allow air to enter the fluid line, which may result either a failure to dispense or an incomplete dispense of the bulk solution on Alinity i and Alinity c. Instructions included that if using Alinity i, and you receive one of the following message codes: 1043, 1044, 1072, 1402, 1403. Refer to Appendix B in the letter. If using Alinity c, and you receive one of the following message codes: 3687, 3689. Refer to Appendix B in the letter. For both Alinity i and c, and those "if" scenarios occur, then discontinue reporting results until troubleshooting is complete and inspect the Alinity ciseries bulk solution level sensors per the instructions in Appendix A in the letter. If a bulk solution level sensor is found to be cracked: Replace the part before running the system using the replacement instructions found in the Alinity ciseries Operations Manual and Abbott recommends following your laboratory procedure to review results produced since your last successful Quality Control (QC). Last, if the level sensor is not cracked, refer to the Alinity ciseries Operations Manual to troubleshoot for the specific message code. To assist in the detection of a cracked bulk solution level sensor, Abbott recommends running Quality Control (QC) immediately before performing weekly maintenance. If your laboratory had previously identified cracks on the bulk solution level sensor before receipt of this communication, and there are messages associated with patient results in the Alinity ciseries System Logs during the same time period, Abbott recommends following your laboratory procedure to review results produced from the replacement of the bulk solution level sensor to the previous successful Quality Co

Details

Recall number
Z-0544-2020
Initiated
September 16, 2019
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
3,271 units
Distribution
Nationwide distribution to AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OK, PR, SC, SD, TN, TX, UT, VA, WA, WI. International distribution to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA…
Recalling firm
Abbott Gmbh & Co. KG, Wiesbaden

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0543-2020 Class IIAlinity ci Level Sensor, Bulk SolutionPotential reliability issue with the Alinity ci series Level Sensor, Bulk Solution. The development of… 09/16/2019Open, Classified
Z-0545-2020 Class IIAccessory kit, Alinity c containing Alinity ci Level Sensor, Bulk SolutionPotential reliability issue with the Alinity ci series Level Sensor, Bulk Solution. The development of… 09/16/2019Open, Classified