Recall Z-0550-2022

Abiomed, Inc. · initiated December 14, 2021

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Codes / lots: UDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04

Reason for recall

The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.

FDA-determined cause: Software design (manufacturing process)

Action taken

The firm began notifying customers by phone starting December 14, 2021. The firm followed up by sending Urgent Medical Device Recall letters to customers on January 20, 2022. The firm is requesting that customers return their devices. The firm does not recommend placing new patients on support. For patients currently on support with an affected system, the firm recommends physicians use their clinical judgement on whether it is appropriate to continue a patient on the OXY-1 System or to transition that patient to the next step in their clinical journey. The letter included instructions on how to return impacted product. If you have questions on returning product, If you have any questions, please contact Diane Welsh from Abiomed s customer service team at (978) 646-1467 or email customerservice@abiomed.com. The firm's Clinical field staff will be available to coordinate boxing and shipping of the OXY-1 system for return to Abiomed. If you have questions regarding this recall, please contact Ralph Barisano directly at (978) 882-8068.

Details

Recall number
Z-0550-2022
Initiated
December 14, 2021
Posted by FDA
February 3, 2022
Terminated
February 28, 2023
Product code
DTZ – Oxygenator, Cardiopulmonary Bypass
Quantity
42 Units
Distribution
US distribution to FL, GA, IN, MD, NC, PA, TX.
Recalling firm
Abiomed, Inc., Danvers, MA