Recall Z-0554-2018
Mindray DS USA, Inc. dba Mindray North America · initiated July 14, 2017
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Codes / lots: All battery chargers with part number 115-030108-00
Reason for recall
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
FDA-determined cause: Component design/selection
Action taken
Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.
Details
- Recall number
- Z-0554-2018
- Initiated
- July 14, 2017
- Terminated
- December 17, 2018
- Product code
- MSX – System, Network And Communication, Physiological Monitors
- Quantity
- 203
- Distribution
- Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
- Recalling firm
- Mindray DS USA, Inc. dba Mindray North America, Mahwah, NJ