Recall Z-0554-2025

Shimadzu Corp. Analytical & Measuring Instruments Div. · initiated October 9, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Codes / lots: Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.

Reason for recall

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

FDA-determined cause: Device Design

Action taken

On 10/9/24, field safety notices were mailed to customers asking them to do the following: Check the Serial Numbers, and if the instruments are affected by this issue, contact your regional office. The corrective actions to be taken to affective products are (1) Add a cover part to the main board. (2) Update the SIL-40 firmware to Ver. 1.16 or above. For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.

Details

Recall number
Z-0554-2025
Initiated
October 9, 2024
Posted by FDA
November 26, 2024
Product code
KIE – Apparatus, High Pressure Liquid Chromatography
Quantity
48
Distribution
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.