Recall Z-0567-2024
Physio-Control, Inc. · initiated November 6, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
Codes / lots: UDI-DI: 00883873940204, Adapters manufactured between 11/3/2021 - 4/24/2023
Reason for recall
AC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.
FDA-determined cause: Component change control
Action taken
On 11/06/23, Recall and Field Safety Alert notices were sent to customers. The recall notices asked customers to do the following: 1) Complete and return business reply form by email to RSRecall@stryker.com 2) Firm recommends continued use. Follow the OI recommendations for daily inspection and testing according to the checklist provided in the Operating Instructions. a. The checklist includes disconnecting the defibrillator from the power adapter, turning the device on, and confirming that two batteries are installed and charged. Then, the defibrillator should be reconnected to the power adapter and users should confirm that the battery charging LED on the defibrillator is illuminated or flashing. b. If an ACPA is experiencing this failure mode, this will present to the user as the Aux Power Indicator on the LP15 flashing instead of solidly illuminating and the green LED strip on the ACPA flashing instead of solidly illuminating: 3) Please notify impacted customers that were shipped or may have been shipped this affected product by sharing this recall notification letter; Email Stryker at RSRecall@stryker.com with a list of customers who received/may have received this product and we can work with you on notifications to impacted customers. If you have any questions or feel that your ACPA may be experiencing issues, please contact Customer Service at 1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday - Friday or by email at medtechsup@stryker.com
Details
- Recall number
- Z-0567-2024
- Initiated
- November 6, 2023
- Posted by FDA
- December 15, 2023
- Product code
- MKJ – Automated External Defibrillators (Non-Wearable)
- Quantity
- 18,732
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
- Recalling firm
- Physio-Control, Inc., Redmond, WA