Recall Z-0568-2026

Meridian Bioscience Inc · initiated October 7, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Meridian Bioscience, alethia c. Difficile, Catalog Number 480050, Nucleic Acid Amplification Test Assay

Codes / lots: UDI/DI 00840733102172, Lot Numbers: 480050U036, Exp. 2026-08-01; 480050U037, Exp. 2026-09-04

Reason for recall

The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.

FDA-determined cause: Nonconforming Material/Component

Action taken

Meridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following: "Required Actions: " Identify remaining inventory of Alethia C. difficile, Lot 480050U036. " Identify the Reaction Buffer bag located within the kit box. " Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket. " Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIRMATION OF NOTIFICATION Form. " Discard Reaction Buffer Tubes which have no gasket present in accordance with local, state, or federal regulations. Note: Execution of the test procedure requires the use of a Reaction Buffer tube. The remaining components in this kit may continue to be utilized for patient testing so long as a new Reaction Buffer tube is available. Once the Reaction Buffer is depleted, the remaining kit components must be discarded in accordance with local regulations. " Meridian Bioscience Inc. will provide credit for the number of tests impacted. " Complete and return the CONFIRMATION OF NOTIFICATION Form to: Product Support Manager, Meridian Bioscience, Inc., 3471 River Hills Drive, Cincinnati, OH 45244; Fax: (513) 272-5432 or Email: FieldActionSupport@meridianbioscience.com. Complete this form if there is no remaining inventory." For any question regarding this notification please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at FieldActionSupport@meridianbioscience.com.

Details

Recall number
Z-0568-2026
Initiated
October 7, 2025
Posted by FDA
November 19, 2025
Product code
OMN – C. Difficile Nucleic Acid Amplification Test Assay
Quantity
109 in total
Distribution
US Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama.
Recalling firm
Meridian Bioscience Inc, Cincinnati, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0569-2026 Class IIMeridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid…The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for… 10/07/2025Open, Classified