Recall Z-0569-2021

Ra Medical Systems, Inc. · initiated October 16, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Codes / lots: Model Number RA-308 Product/Catalog Number: 1200-0001-01 Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.

Reason for recall

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

FDA-determined cause: Device Design

Action taken

On 10/29/2020, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification via certified mail informing them of the potential for the footswitch to fail to operate properly if liquid is spilled near or on the footswitch. On 11/16/2020, the firm sent out an updated customer notification. The firm is instructing customers to discontinue using the footswitch if liquids have been spilled on or splashed on the floor near or on the footswitch. The Recalling Firm will ship a replacement footswitch and will provide shipping materials for users to return the old footswitch. Instructions will be provided on how to install the new footswitch and return the old footswitch. Customers are asked to fill out and return the Acknowledgement and Receipt Form via FAX to 760-867-2706 or email at fieldcorrectiveaction@ramed.com. For any questions, contact the Recalling Firm at: 1-877-635-1800 Monday thru Friday 8:00 a.m. to 5:00 p.m. PST

Details

Recall number
Z-0569-2021
Initiated
October 16, 2020
Terminated
January 11, 2023
Product code
PDU – Catheter For Crossing Total Occlusions
Quantity
36 devices
Distribution
US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.
Recalling firm
Ra Medical Systems, Inc., Carlsbad, CA

Official FDA recall record