Recall Z-0583-2011

Teco Diagnostics · initiated October 1, 2010

Terminated

Product

NDC, Inc. Pro Advantage Urine Analyzer, Catalog Number: Pro Advantage Urine Analyzer The Uritek TC-101 (TC-101) urine chemistry analyzer is a semi-automated analyzer for urine chemistry analysis. This device has been manufactured and developed with advanced scientific technology of modern microprocessors, optics and electronics. The apparatus uses advanced testing technology of a cold-light source, integrating sphere, and modularized blocking systematical design which enhances the stability of the optical system. This system also automatically compensates for the influence of ambient and scattered lights through the coordination of software and hardware. The TC-101 can test for 10 parameters including Glucose (GLU), Urobilinogen (URO), Bilirubin (BIL), Ketone (KET), Specific Gravity (SG), Occult Blood (BLD), pH, Protein (PRO), Nitrite (NIT), Leukocytes (LEU) in 60 seconds when used with Urine Reagent Strips (URS) 10.

Codes / lots: All units sold (2005 - 2010)

Reason for recall

This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for use in measuring urinary glucose or occult blood with the CLIA waived, Inc. CLIA-10 test strips. Teco Diagnostics has been notified by the FDA that the CLIA-101 requires 510(k) clearance and is not CLIA Waived for use with the CLIA-10 test strips. Without FDA approval or clearance, Teco Di

FDA-determined cause: Other

Action taken

Teco Diagnostics initiated an "Urgent: Medical Device Recall" letter dated October 14, 2010, with an attached "Recall Return Response Form" to the affected consignees via US Postal on October 15, 2010. The recall notice informed the consignees of the reason for recall, the products affected and the actions required. Consignees were instructed to complete and return the enclosed Recall Return Response Form as soon as possible via mail, fax, or email. Upon receipt of the Recall Return Response Form, Teco Diagnostics would work with the consignees/distributors to process any refunds or reimbursements for the returned product and return shipping. If consignees/distributor had questions, they were to call Teco's Technical Support Department at (800) 222-9880 or (714) 463-1111.

Details

Recall number
Z-0583-2011
Initiated
October 1, 2010
Posted by FDA
December 9, 2010
Terminated
August 3, 2012
Product code
KQO – Automated Urinalysis System
Quantity
2,140 units total for all products
Distribution
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, AND WV and the country of Canada.
Recalling firm
Teco Diagnostics, Anaheim, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0577-2011 Teco Diagnostics Uritek TC-101 Urine Analyzer, Catalog Number: TC-101 The Uritek TC-101…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated
Z-0578-2011 Cliawaived, Inc. CLIA-101 Urine Analyzer, Catalog Number: CLIA-101 The Uritek TC-101…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated
Z-0579-2011 Cole Taylor Marketing, Inc. CT-101 Urine Analyzer, Catalog Number: CT-101 The Uritek…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated
Z-0580-2011 Henry Schein, Inc. OneStep Pro Urine Analyzer, Catalog Number: OneStep Pro Urine…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated
Z-0581-2011 Immunostics, Inc. Detector UriStrip Urine Analyzer, Catalog Number: Detector UriStrip…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated
Z-0582-2011 McKesson Medical Surgical 107-101 Urine Analyzer, Catalog Number: 107-101 The Uritek…This recall has been initiated due to the CLIA-101 Urine Analyzer not having 510(k) clearance by the FDA for… 10/01/2010Terminated