Recall Z-0601-2014

Asante Solutions, Inc. · initiated November 19, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Asante Conset Infusion Set with tubing & 9mm cannula packaged in 5-packs,singles,or inside a Combination Kit (FG-5249-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.

Codes / lots: Asante Catalog number:FG-4249-5, FG-4249T; Unomedical Item number: 43-110-5269; Lot numbers: 0230129 5005804 5040412.

Reason for recall

Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.

FDA-determined cause: Device Design

Action taken

Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.

Details

Recall number
Z-0601-2014
Initiated
November 19, 2013
Posted by FDA
December 27, 2013
Terminated
April 24, 2014
Product code
FPA – Set, Administration, Intravascular
Quantity
1085 kits
Distribution
Nationwide Distribution.
Recalling firm
Asante Solutions, Inc., Sunnyvale, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0594-2014 Class IIAsante Comfort Infusion Set with 60cm tubing & 13mm cannula packaged in 5-packs…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0595-2014 Class IIAsante Comfort Infusion Set with 60cm tubing & 17mm cannula packaged in 5-packs…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0596-2014 Class IIAsante Comfort Infusion Set with 110cm tubing & 13mm cannula packaged in…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0597-2014 Class IIAsante Comfort Infusion Set with 110cm tubing & 17mm cannula packaged in 5-packs…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0598-2014 Class IIAsante Conset Infusion Set with 60cm tubing & 6mm cannula packaged in 5-packs,singles,or…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0599-2014 Class IIAsante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated
Z-0600-2014 Class IIAsante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs…Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the… 11/19/2013Terminated