Recall Z-0602-2021
Cook Inc. · initiated November 24, 2020
Class I – dangerous or defective products that could cause serious health problems or death
Product
Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL0-HC, Order Number G56232 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.
Codes / lots: Lot Numbers: 13205314, 13205322, 13229673, 13229685, 13234800, 13249450, 13256702, 13262338, 13265332, 13265337, 13279097, 13288121, 13291877, 13297249, 13297260, 13297261, 13301872, 13310761, 13312706, 13318040, 13318062, 13332170, 13346105, 13349029, 13354271, 13354272, 13357259, 13360427, 13360442, 13366230, 13372465, 13372469, 13375211, 13375227, 13377002, 13391151, 13398856, 13426338, 13329254X, 13332169X, 13342871X
Reason for recall
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
FDA-determined cause: Process control
Action taken
The firm, Cook Medical, sent "URGENT: MEDICAL DEVICE RECALL" letters dated 11/24/20 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This medical device recall letter must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing devices potentially impacted from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue. Other Information This action is being taken with the knowledge of the Food and Drug Administration. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to
Details
- Recall number
- Z-0602-2021
- Initiated
- November 24, 2020
- Posted by FDA
- December 23, 2020
- Terminated
- September 7, 2022
- Product code
- DYB – Introducer, Catheter
- Quantity
- 1967
- Distribution
- US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong…
- Recalling firm
- Cook Inc., Bloomington, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0591-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL0-HC, Order Number G56229 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0592-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL1-HC, Order Number G56234 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0593-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-4.0-35-70-RB-HFANL0-HC, Order Number G56219 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0594-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-5.0-35-55-RB-HFANL0-HC, Order Number G56230 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0595-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-5.0-35-55-RB-HFANL1-HC, Order Number G56235 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0596-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-5.0-35-70-RB-HFANL0-HC, Order Number G56220 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0597-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-5.0-35-70-RB-HFANL1-HC, Order Number G56225 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0598-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0599-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0600-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL0-HC, Order Number G56221 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0601-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL1-HC, Order Number G56226 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0603-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL1-HC, Order Number G49043 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0604-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-7.0-35-70-RB-HFANL0-HC, Order Number G56222 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0605-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-7.0-35-70-RB-HFANL1-HC, Order Number G56227 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0606-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0607-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL1-HC, Order Number G49045 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0608-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL0-HC, Order Number G56223 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0609-2021 Class I | Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL1-HC, Order Number G56228 Flexor…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0610-2021 Class I | Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |
| Z-0611-2021 Class I | Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN…There is an increased likelihood of the introducer sheath separating at the proximal bond site. If… | 11/24/2020Terminated |