Recall Z-0604-2013

Medrad Inc · initiated April 18, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medrad(R) Continuum MR Infusion System Secondary VSA Set Peristaltic Pump System designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR procedure functioning while either stationary or mobile.

Codes / lots: Catalog number - Secondary VSA Set: 3015161. See firm's website - www.medrad.com for specific serial and lot numbers.

Reason for recall

Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and microbore), catalog numbers MIK200A, MIL 200B, 3015153, 3015155, 3015156, 3015157, 3015158, 3015159, and 3015161. The firm is initiating this recall because this tubing exhibits a variation that, when used in conjunction with any Continuum pump, may result in flow rate accuracy performance issues.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm, Bayer HealthCare, sent an "URGENT MEDICAL DEVICE RECALL" letter dated April 17, 2012 via the USPS certified mail to its consignees/customers. The firm expanded the recall on May 11, 2012 and sent a second letter entitled "EXPANDED URGENT MEDICAL DEVICE RECALL" to its affected consignees/customers notifying them of the additional issue. The letters described the product, problem and actions to be taken. The customers were instructed to return pump now; contact customer support at 1-877-229-3767 for a return goods authorization (RGA); complete and return Response Form, include RGA numbers, via Fax to Bayer Compliance at 412-4060942, and/or wait to return the pump when they begin contacting customers with this subset of pumps in August to make arrangements for a replacement pump. If you have questions, please contact our customer support team at 1-877-229-3767 or visit www.medrad.com for more information.

Details

Recall number
Z-0604-2013
Initiated
April 18, 2012
Posted by FDA
December 20, 2012
Terminated
August 17, 2015
Product code
FPA – Set, Administration, Intravascular
Quantity
1,229 Pumps total
Distribution
Worldwide distribution: USA (nationwide) and countries of: Australia, Bermuda, Brazil, Canada, China, Chile, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Great Britain, Guam, Hong Kong, India, Israel, Italy, Japan, Korea, Lebanon, Mexico, Malaysia, Netherland, Poland, Portugal, Puerto Rico, Sweden, Singapore, Taiwan, Thailand, and Venezula.
Recalling firm
Medrad Inc, Indianola, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0595-2013 Class IIMedrad(R) Continuum MR Infusion Pump Peristaltic Pump System designed for patients who…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0596-2013 Class IIMedrad(R) Continuum MR Infusion System Standard Administration Kit Peristaltic Pump…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0597-2013 Class IIMedrad(R) Continuum MR Infusion System Secondary Administration Kit Peristaltic Pump…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0598-2013 Class IIMedrad(R) Continuum MR Infusion System Primary Set Peristaltic Pump System designed for…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0599-2013 Class IIMedrad(R) Continuum MR Infusion System Primary VSA Stopcock Set Peristaltic Pump System…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0600-2013 Class IIMedrad(R) Continuum MR Infusion System Primary VSA Set Peristaltic Pump System designed…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0601-2013 Class IIMedrad(R) Continuum MR Infusion System Primary Spike Set Peristaltic Pump System…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0602-2013 Class IIMedrad(R) Continuum MR Infusion System Secondary Set Peristaltic Pump System designed…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0603-2013 Class IIMedrad(R) Continuum MR Infusion System Secondary Spike Set Peristaltic Pump System…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated
Z-0605-2013 Class IIMedrad(R) Continuum MR Infusion System Side Step Set Peristaltic Pump System designed…Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets (macrobore and… 04/18/2012Terminated