Recall Z-0620-2011

Spacelabs Healthcare, Llc · initiated March 20, 2009

Terminated

Product

Spacelabs Medical Flat Panel Display is a touchscreen display, 19-inch thin film transistor liquid crystal display (TFT-LCD). The PNs are 010-1619-00 and 010-1655-00, Models 94260-19 and 94256.

Codes / lots: Date Codes (WK numbers) identified on the affected power supplies: 0647, 0703, 0706, 0707, 0709, 0712, 0724, 0727, 0728 or 0732. Lot/Serial Numbers on the systems for US Consignees: ELZ-0011442, ELZ-0012143, ELZ-0013307, ELZ-0013557, ELZ-0014567, ELZ-0014570, ELZ-0014574, ELZ-0014577, ELZ-0014580, ELZ-0014583, ELZ-0014585, ELZ-0014586, ELZ-0014587, ELZ-0014590, ELZ-0014591, ELZ-0014592, ELZ-0014593, ELZ-0014594, ELZ-0014595, ELZ-0014596, ELZ-0014597, ELZ-0014598, ELZ-0014599, ELZ-0014600, ELZ-0014601, ELZ-0014603, ELZ-0014604, ELZ-0014605, ELZ-0014606, ELZ-0014607, ELZ-0014608, ELZ-0014609, ELZ-0014610, ELZ-0014611, ELZ-0014612, ELZ-0014613, ELZ-0014614, ELZ-0014615, ELZ-0014616, ELZ-0014617, ELZ-0014618, ELZ-0014619, ELZ-0014620, ELZ-0014621, ELZ-0014622, ELZ-0014623, ELZ-0014625, ELZ-0014626, ELZ-0014629, ELZ-0014630, ELZ-0014631, ELZ-0014632, ELZ-0014633, ELZ-0014634, ELZ-0014635, ELZ-0014636, ELZ-0014637, ELZ-0014638, ELZ-0014639, ELZ-0014640, ELZ-0014641, ELZ-0014645, ELZ-0014648, ELZ-0014656, ELZ-0014657, ELZ-0014658, ELZ-0014718, ELZ-0014719, ELZ-0014720, ELZ-0014723, ELZ-0014724, ELZ-0014726, ELZ-0014727, ELZ-0014729, ELZ-0014731, ELZ-0014733, ELZ-0014736, ELZ-0014737…

Reason for recall

The power supply of the Spacelabs Medical Flat Panel Display may fail, causing the display to go blank.

FDA-determined cause: Other

Action taken

Spacelabs Healthcare sent an URGENT: MEDICAL DEVICE RECALL Spacelabs Medical Flat Panel Power Supply PNs: 010-1619-00 & 010-1655-00 letter dated March 20, 2009 to US consignees. On March 27, 2009, Spacelabs sent the same recall letter with translation option to their international consignees.The letter stated the firm would be sending a replacement power supply. While waiting for the replacement of the power supply, Spacelabs Medical recommended their consignees brief the staff on the possibility of the display going blank to help prevent any adverse event from happening. The consignees were advised to dispose the old power supply per their premise protocol or local regulations. The consignees from US can contact Spacelabs Medical at 1-800-522-7025, #2 for Technical Support , and consignees outside the United States can call 1-425-657-7200 x5089 for any questions about the recall.

Details

Recall number
Z-0620-2011
Initiated
March 20, 2009
Posted by FDA
December 14, 2010
Terminated
June 13, 2011
Product code
DXJ – Display, Cathode-Ray Tube, Medical
Quantity
2877 units in US and 731 units outside US
Distribution
Worldwide Distribution: USA in the following states: AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, PA, RI, SC, SD, TN, TX, VA, WA, WI, and WY, and including the following countries: AUSTRALIA, AUSTRIA, CANADA, Caribbean, CZECH REPUBLIC, EGYPT, ESTONIA, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, ISRAEL…
Recalling firm
Spacelabs Healthcare, Llc, Issaquah, WA

Official FDA recall record