Recall Z-0621-2025

Becton Dickinson & Co. · initiated November 6, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD BBL Sensi Disc Ampicillin 2 g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263

Codes / lots: UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).

Reason for recall

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

FDA-determined cause: Under Investigation by firm

Action taken

BD issued URGENT: Medical Device Product Correction (IDS-24-5142) letter on 11/6/24 to Distributors (included Customer Letter) via e-mail, Letter states reason for recall, health risk and action to take: Actions for Clinical Users: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood by those within your organization. 3. There are no additional recommendations for repeat testing or review of prior results. 4. Share and post this customer letter with users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 5.Complete the attached Customer Response Form and return to your distributor, whether or not you have any of the impacted material. If you require further assistance, please contact: BD Contact Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com Product Complaints, Technical Questions Post-Market Quality BDRC42@bd.com Recall Questions

Details

Recall number
Z-0621-2025
Initiated
November 6, 2024
Posted by FDA
December 5, 2024
Product code
JTN – Susceptibility Test Discs, Antimicrobial
Quantity
13,190 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.
Recalling firm
Becton Dickinson & Co., Sparks, MD

Official FDA recall record