Recall Z-0625-2014

Siemens Healthcare Diagnostics, Inc. · initiated April 22, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

HB1C Flex(R) Reagent Cartridge (DF105A, Siemens Material Number 10483822) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R) reagent Cartridge on the Dimension(R) clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.

Codes / lots: Catalog number DF105A (Siemens Material Number 10483822) - Lot numbers GA3099, exp. 4/9/13; GA3141, exp. 5/21/13; GA3197, exp. 7/16/13; GA3247, exp. 9/4/13; GA3113, exp. 4/23/13; GA3162, exp. 6/11/13; GB3211, exp. 7/30/13; GA3267, exp. 9/24/13; GA3134, exp. 5/14/13; GA3169, exp. 6/18/13; GA3232, exp. 8/20/13; GB3281, exp. 10/8/13, GC3302, exp. 10/29/13.

Reason for recall

Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.

FDA-determined cause: Process control

Action taken

Siemens sent an Urgent Medical Device Recall letter in April 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discard any remaining inventory of the affected product. Siemens will replace any unused inventory at no charge. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate they have received the information. Customers were asked to retain the letter with their laboratory records and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local Siemens technical support representative. For questions regarding this recall call 1-800-441-9250.

Details

Recall number
Z-0625-2014
Initiated
April 22, 2013
Posted by FDA
January 3, 2014
Terminated
February 12, 2014
Product code
LCP – Assay, Glycosylated Hemoglobin
Quantity
11006 Total
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0626-2014 Class IIHbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) for the…Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range"… 04/22/2013Terminated