Recall Z-0629-2018

Cincinnati Sub-Zero Products LLC · initiated November 27, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Hemotherm Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.

Codes / lots: none

Reason for recall

Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

FDA-determined cause: Device Design

Action taken

Between November 27, 2017 - December 8, 2017, Cincinnati Sub-Zero Products LLC, distributed Urgent Medical Device Correction notices to their customers via Email and USPS. This Medical Device Correction has been initiated to provide a labeling update to the cleaning & disinfection procedures. The update addresses the disinfection process, cleaning of reusable accessories, as well as clarifying the water type that should be used with the Hemotherm. Customers are encouraged to do the following: 1) Discard your current Hemotherm Operation & Technical Manual and/or Operation Manual. 2) Replace your current Hemotherm Operation & Technical Manual and/or Operation Manual with the enclosed updated Manual(s). For your convenience, all updates have been highlighted in the manuals. " For Hemotherm 400CE users, utilize the P/N 56418-R Operation & Technical Manual and P/N 56075-G Operation Manuals. " For Hemotherm 400MR users, utilize the P/N 56234-S Operation Manual. All customers are encouraged to complete and return the enclosed response form by fax or email to FA2016-009@genthermcsz.com; Fax: 513-772-9119. Customers with questions or concerns, please contact the Field Action Coordinator: Kathy DeSmidt, E-mail: FA2016-009@genthermcsz.com, Phone: 513-772-8810 or 800-989-7373.

Details

Recall number
Z-0629-2018
Initiated
November 27, 2017
Posted by FDA
February 13, 2018
Terminated
July 2, 2018
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
1355 units
Distribution
US Nationwide Distribution
Recalling firm
Cincinnati Sub-Zero Products LLC, Cincinnati, OH