Recall Z-0644-2015
Stryker Endoscopy · initiated November 17, 2014
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Codes / lots: 140280 143823 244698 341493 443333 541776 545061 647416 744587 842092 844289 846109 846772 846819 941064 941065 941852 942140 943501 943549 944300 945744 945886 945977 1040098 1040413 1041057 1041621 1041992 1042086 1042240 1042475 1042685 1042695 1042698 1042714 1044482 1045035 1045373 1045641 1046199 1046595 1046933 1046934 1046943 1142155 1142668 1142897 1142910 1143132 1144208 1144398 1144492 1144940 1145080 1145798 1145971 1146839 1146933 1240119 1241514 1241515 1242921 1244191 1244534 1245445 1246192E 1246638 1246638E 1246650 1340099 1341584 1341585 1343125 1343402 1343402E 1343705 1344204 1344286 1344456 1344880 1345236 1345532 1345532E 1345605 1345605E 1346432 1346666 1440405 1440684 1442518 1443219 1443653 1443917
Reason for recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
FDA-determined cause: Error in labeling
Action taken
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Details
- Recall number
- Z-0644-2015
- Initiated
- November 17, 2014
- Posted by FDA
- December 12, 2014
- Terminated
- July 17, 2015
- Quantity
- 180,573 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South…
- Recalling firm
- Stryker Endoscopy, San Jose, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0592-2015 Class II | PKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide:…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0593-2015 Class II | PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080120 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0594-2015 Class II | PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0595-2015 Class II | PKG, BIPOLAR FORCEPS, MICRO TIP, 33CM, P/N 0250080122 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0596-2015 Class II | PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0597-2015 Class II | PKG, OP-BIOPSY FORCEPS, SPOON JAW, P/N 0250080007 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0598-2015 Class II | PKG, FIXATION CLAMP 5MM WITH RATCHET, P/N 0250080130 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0599-2015 Class II | PKG, KNOT PUSHER 5MM OPEN, P/N 0250080132 Laparoscopic Manual Instruments are intended…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0600-2015 Class II | PKG, MYOMA DRILL 5MM LENGTH 320MM, P/N 0250080133 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0601-2015 Class II | PKG, PALPATION PROBE GRADUATED 5MM, P/N 0250080135; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0602-2015 Class II | PKG, PORTIO ACORN SMALL ONLY, P/N 0250080139; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0603-2015 Class II | PKG, TROCAR 5.5MM PYRAMIDAL TIP, P/N 0250080154; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0604-2015 Class II | PKG, OBTURATOR, 5.5MM FOR HASSON CANNULA, P/N 0250080157; Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0605-2015 Class II | PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0606-2015 Class II | PKG, MARYLAND FORCEPS, CURVED, P/N 0250080009. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0607-2015 Class II | PKG, HASSON CANNULA, 5.5MM, AUTOMATIC VALVE, STOP COCK, P/N 0250080176. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0608-2015 Class II | PKG, MIXTER FORCEPS, P/N 0250080010. Laparoscopic Manual Instruments are intended for…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0609-2015 Class II | PKG, HOOK SCISSORS, P/N 0250080020. Laparoscopic Manual Instruments are intended for…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0610-2015 Class II | PKG, HASSON CANNULA, 11MM AUTOMATIC VALVE, STOP COCK, P/N 0250080179. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0611-2015 Class II | PKG, FIXABLE CONE FOR 11MM HASSON, P/N 0250080180. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0612-2015 Class II | PKG, METZENBAUM SCISSORS, STRAIGHT, P/N 0250080024. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0613-2015 Class II | PKG, METZENBAUM SCISSORS, CURVED, LONG, P/N 0250080025. Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0614-2015 Class II | PKG, BABY METZENBAUM SCISSORS, CURVED, P/N 0250080028. Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0615-2015 Class II | PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080029. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0616-2015 Class II | PKG, 12.5/11MM - 5.5 REDUCER, P/N 0250080204. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0617-2015 Class II | PKG, BULLET FORCEPS, TAPERED, SERRATED, SPOON, P/N 0250080041. Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0618-2015 Class II | PKG, LUERLOCK, MALE, 6MM TUBING, P/N 0250080207. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0619-2015 Class II | PKG, LUERLOCK, FEMALE, 6MM TUBING, P/N 0250080208. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0620-2015 Class II | PKG, MARYLAND FORCEPS, CURVED, LEFT, P/N 0250080042. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0621-2015 Class II | PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080230. Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |