Recall Z-0644-2018

Innovasis, Inc · initiated January 8, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Excella II Polyaxial Pedicle Screw used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:

Codes / lots: Catalog/Part No.: E2S8530, E2S8535, E2S8540, E2S8545, E2S8550, E2S8555, E2S8560 with Lot Number: 1101 and Catalog/Part No.: E28540, E28545 with Lot Number: 1201. Each of these parts is engraved with part number and lot number.

Reason for recall

Nine lots of 8.5 mm Polyaxial Pedicle Screws are at risk of breakage.

FDA-determined cause: Nonconforming Material/Component

Action taken

The distributors were notified of the recall on 1/8/2018 via email with an attached pdf. They are also being asked to confirm or deny the quantities showing in inventory via a Product Removal Form sent in the same email.

Details

Recall number
Z-0644-2018
Initiated
January 8, 2018
Terminated
February 27, 2018
Product code
NKB – Thoracolumbosacral Pedicle Screw System
Quantity
1,010 parts
Distribution
United States
Recalling firm
Innovasis, Inc, Salt Lake City, UT

Official FDA recall record