Recall Z-0646-2024

Abiomed, Inc. · initiated November 7, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Abiomed Axillary Insertion Introducer -Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3006 Individual, and Part of Pump Set (005062, 0550-0008, 1000100, 0052-0011, 1000220)

Codes / lots: UDI-DI: 00885672009755

Reason for recall

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

FDA-determined cause: Process control

Action taken

Abiomed issued URGENT MEDICAL DEVICE RECALL (REMOVAL) Letter Products Impacted: Introducer Kit for Impella (sold within Impella Pump sets and Individually Packaged) Axillary Introducer Kits on 11/7/23. Letter states reason for recall, health risk and action to take: IF AN IMPELLA PUMP IS CLINICALLY NEEDED, ACTIONS TO BE TAKEN BY THE USER: 1. Prior to usage, carefully examine the outer pouch of the Introducer Kit for holes following the provided instructions in Attachment 2. 2. For non-trayed introducers (Part Numbers 0052-3006): IF THERE IS ANY SUSPICION OF THIS TYPE OF PACKAGING DAMAGE, REFRAIN FROM USING THE PRODUCT. 3. For trayed introducers with an inner tray package (Part Numbers 0052-3014, 0052-3015, 0052-3021, 0052-3052, 0052-3046, 0052-3025, 0052-3056, 2000342, 0052-3053): Consider that while the outside surface of the tray is non-sterile, the introducers inside the tray are considered sterile. After removing the inner tray from the outer pouch, avoid introducing it to the sterile field. Instead, employ sterile technique to open the inner tray, preserving its interior sterility. a. Transfer the contents from the inner tray to the sterile field following your institution s sterile field management protocol, ensuring no contact with the outer surface to prevent contamination. b. The inner tray must remain separate from the sterile field and should be discarded once all components of the introducer kit have been extracted. 4. If used, document the use on the Business Request Form at the end of this letter. Please contact Abiomed s Clinical Support Center, 1-800-422-8666, if you need any support troubleshooting or if you have any further questions.

Details

Recall number
Z-0646-2024
Initiated
November 7, 2023
Posted by FDA
January 5, 2024
Product code
DYB – Introducer, Catheter
Quantity
717 single units; 1906 within sets
Distribution
Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0647-2024 Class IIOscor Introducer Kit, 14Fr x Short Individual (0052-0001)- Introducer(s) are intended…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0648-2024 Class IIAbiomed Oscor Introducer Kit, 14Fr x Long Individual (0052-0017)- Introducer(s) are…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0649-2024 Class IIAbiomed Introducer Kit, 23 Fr, w/Hlx Vlv Sterile Individual, and Part of Pump Set…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0650-2024 Class IIAbiomed Kit, 14Fr Introducer, 13cm, Long Taper Individual (0052-0038)-Introducer(s) are…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0651-2024 Class IIAbiomed Kit, 14Fr Introducer, 25cm, Long Taper Individual (0052-0039)-Introducer(s) are…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0652-2024 Class IIAbiomed Kit, 14Fr Introducer, 13cm&25cm, Sterile Part of Pump Set (0048-0040, 0048-0045…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0653-2024 Class IIOscor Intro Kit, 14Fr x 13 and 25 cm Part of Pump Set (0048-0040)-Introducer(s) are…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0654-2024 Class IIAbiomed KIT, 14 Fr Introducer, 13 & 25 cm, ABMD Part of Pump Set (0048-0024…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified
Z-0655-2024 Class IIAbiomed Kit, 23Fr Introducer, 11cm, Sterile Part of Pump Set (1000323)-Introducer(s) are…Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile… 11/07/2023Open, Classified