Recall Z-0647-2010

Spacelabs Healthcare, Incorporated · initiated November 6, 2009

Terminated

Product

Spacelabs Model 91367 SL 2200 Compact Monitor Spacelabs Medical patient monitors passively display data generated from Spacelabs Medical patient parameter modules, Flexport interfaces, gas analyzers, and other SDLC based products.

Codes / lots: S/N, U.S. Domestic units: 1367-000395, 1367-000396, 1367-000397, 1367-000628, 1367-000629, 1367-000630, 1367-000631, 1367-000632, 1367-000633, 1367-000522, 1367-000525, 1367-000526, 1367-000527, 1367-000601, 1367-000602, 1367-000603, 1367-000604, 1367-000605, 1367-000606, 1367-000477, 1367-000731, 1367-000732, 1367-000733, 1367-000734, 1367-000444, 1367-000831, 1367-000832, 1367-000833, 1367-000437, 1367-000438, 1367-000272, 1367-000454, 1367-000482, 1367-000483, 1367-000433, 1367-000434, 1367-000435, 1367-000436, 1367-000479, 1367-000480, 1367-000481, 1367-000524, 1367-000399, 1367-000400, 1367-000411, 1367-000412, 1367-000735, 1367-000736, 1367-000737, 1367-000738, 1367-000739, 1367-000740, 1367-000741, 1367-000742, 1367-000743, 1367-000744, 1367-000745, 1367-000746, 1367-000747, 1367-000748, 1367-000749, 1367-000439, 1367-000440, 1367-000441, 1367-000442, 1367-000443, 1367-000523, 1367-000530, 1367-000531, 1367-000532, 1367-000663, 1367-000664, 1367-000665, 1367-000691, 1367-000692, 1367-000693, 1367-000694, 1367-000695, 1367-000696, 1367-000697, 1367-000698, 1367-000699, 1367-000700, 1367-000701, 1367-000702, 1367-000703, 1367-000704, 1367-000705, 1367-000706, 1367-000707…

Reason for recall

Potential for monitor to shut down and not being able to power on again.

FDA-determined cause: Component design/selection

Action taken

On 11/6/09 the firm sent Urgent - Device Correction notifications to customers by certified mail informing them of the monitor failure and that a field service engineer will contact the customer to fix the affected device at no cost. On 11/13/2009 the firm's Global Tech Support sent notification via email to all foreign subsidiaries and distributors informing them of the monitor failure and that a field service engineer will contact the customer to fix the affected device at no cost. Questions are to be directed to Spacelabs Medical's Technical Support at 1-800-522-7025, and 1-425-657-7200 x5089 for those customers outside of the US.

Details

Recall number
Z-0647-2010
Initiated
November 6, 2009
Posted by FDA
January 19, 2010
Terminated
July 21, 2011
Product code
DXJ – Display, Cathode-Ray Tube, Medical
Quantity
422 units
Distribution
Worldwide Distribution
Recalling firm
Spacelabs Healthcare, Incorporated, Issaquah, WA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0648-2010 Spacelabs Model 91369-24 SL 2400 Compact Monitor Spacelabs Medical patient monitors…Potential for monitor to shut down and not being able to power on again. 11/06/2009Terminated
Z-0649-2010 Spacelabs Model 91370 SL 2600 Compact Monitor Spacelabs Medical patient monitors…Potential for monitor to shut down and not being able to power on again. 11/06/2009Terminated