Recall Z-0650-2014

Nova Biomedical Corporation · initiated October 10, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Codes / lots: All serial numbers

Reason for recall

Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

FDA-determined cause: Component design/selection

Action taken

On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.

Details

Recall number
Z-0650-2014
Initiated
October 10, 2013
Posted by FDA
January 9, 2014
Terminated
March 24, 2014
Product code
CGA – Glucose Oxidase, Glucose
Quantity
est. 200,000
Distribution
Nationwide Distribution.
Recalling firm
Nova Biomedical Corporation, Waltham, MA

Official FDA recall record