Recall Z-0651-2009
Snug Seat Inc · initiated November 6, 2008
Product
High-low:x frame for the x:panda mechanical chair.
Codes / lots: Article numbers 8911605-02, 8911605-02E, 8911606-02, 8911615-02, 8911615-02E and 8911616-02. Manufactured between 8/24/2007 and 10/3/2008.
Reason for recall
Breakage in the welds was discovered in a section of the chair frame.
FDA-determined cause: Process control
Action taken
Consignees were notified by letter dated 11/6/2008. Snug Seat sent a repair kit to their retail dealers on 11/18/2008. The kit includes a safety plate, written and video replacement instructions as well as the tools (allen wrench, crescent wrench and two nuts) necessary to make the safety update. Customers were identified using purchase order numbers, and product serial numbers to enable the dealers to identify the purchaser of the product. Due to HIPPA regulations, purchaser information was not available to Snug Seat, Inc. to contact directly. Written notification will be followed up with a telephone call to each dealer to ensure they have received the retrofit kit. Snug Seat will also insure they understand the importance of the kit and how to properly install and document receipt of the update for their customers.
Details
- Recall number
- Z-0651-2009
- Initiated
- November 6, 2008
- Posted by FDA
- January 5, 2009
- Terminated
- February 17, 2011
- Product code
- IMS – Support, Head And Trunk, Wheelchair
- Quantity
- 67 units
- Distribution
- Nationwide.
- Recalling firm
- Snug Seat Inc, Stallings, NC