Recall Z-0654-2019

Bracco Injeneering S.A. · initiated November 12, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Empower FastLoad CT Syringe Pack -sterile disposable syringe kit for use with the EmpowerCT, EmpowerCTA and EmpowerCTA+ Injector Systems SKU 017344/6720 Empower FastLoad" CT Syringe Pack is a sterile, disposable syringe kit designed for use with EmpowerCT, EmpowerCTA and EmpowerCTA+ Injector Systems. It is used to intravenously facilitate the administration of contrast media and flushing solutions into the human vascular systems.

Codes / lots: Lot # IANBL-1804 UDI: (01) 17630039300309 (17) 210420 (11) 180420 (10) IANBL-1804

Reason for recall

Insufficient seal on the sterile barrier of the device, compromising the sterility of the device

FDA-determined cause: Packaging process control

Action taken

The firm, Bracco Injeneering, issued "URGENT: MEDICAL DEVICE RECALL" letter via third party recall firm, on 11/12/18 to customers and distributors identifying the problem, and requesting return of unused product., Complete and return the Removal Recall Response Form even if you have no recalled stock on hand. Questions: contact Stericycle at 888-759-6907 (Monday- Friday 8 a.m. to 5 p.m. Eastern Time) or acistmedical3021@stericycle.com.

Details

Recall number
Z-0654-2019
Initiated
November 12, 2018
Terminated
April 15, 2020
Product code
DXT – Injector And Syringe, Angiographic
Quantity
33,350 packs (667 cases)
Distribution
US Nationwide Distribution
Recalling firm
Bracco Injeneering S.A., Lausanne