Recall Z-0664-2008

Fertility Technology Resources, Inc. · initiated September 13, 2007

Terminated

Product

Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 20 cm, Order #: ftr120, Manufactured by: Fertility Technology Resources, Inc., Marietta, GA 30062

Codes / lots: All units shipped before August 17, 2007

Reason for recall

Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance.

FDA-determined cause: Process design

Action taken

All customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.

Details

Recall number
Z-0664-2008
Initiated
September 13, 2007
Posted by FDA
January 19, 2008
Terminated
February 6, 2008
Product code
MQF – Catheter, Assisted Reproduction
Quantity
not reported
Distribution
Worldwide: USA including states of CA, CO, FL, GA, LA, MD, ME, MI, MS, NC, NH, NJ, NY, OH, TN, and WA and The Netherlands.
Recalling firm
Fertility Technology Resources, Inc., Marietta, GA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0662-2008 Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 23 CM, Sterile…Expiration Date for Sterility; The 5 year dating could not be validated/supported for sterility assurance. 09/13/2007Terminated
Z-0663-2008 Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 20 cm, Sterile…Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance. 09/13/2007Terminated
Z-0665-2008 Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 23 cm, Order #: ftr123…Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance. 09/13/2007Terminated
Z-0666-2008 Tucker Embryo Catheter for Trial Transfer, 20 cm, Order #: ftr320, Manufactured by:…Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance. 09/13/2007Terminated
Z-0667-2008 Morton IUI Catheter, Sterile, Single Use Only, Order #: ftr510, Manufactured by…Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance. 09/13/2007Terminated