Recall Z-0667-2019

Abbott Ireland Diagnostics Division · initiated November 21, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Alinity i Estradiol Reagent Kit, List Number 07P5020

Codes / lots: UDI/Lot Numbers (01)00380740131159 (17)190317(10) 90193UI00 / 90193UI00

Reason for recall

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm initiated the recall by letter on 11/21/2018. The notice explained the problem and provided instruction on how to inspect the reagent cartridges for defect. The consignee was directed to discard the defective units. The units will be replaced.

Details

Recall number
Z-0667-2019
Initiated
November 21, 2018
Terminated
April 1, 2021
Product code
CHP – Radioimmunoassay, Estradiol
Quantity
79 units
Distribution
AR, FL, GA, MD, OK, VA
Recalling firm
Abbott Ireland Diagnostics Division, Lisnamuck Co. Longford

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0664-2019 Class IIAlinity i TSH Reagent Kit, List Number 07P4830Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the… 11/21/2018Terminated
Z-0665-2019 Class IIAlinity i Free T4 Reagent Kit, List Number 07P7030Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the… 11/21/2018Terminated
Z-0666-2019 Class IIAlinity i Total Beta-hCG Reagent Kit, List Number 07P5121Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the… 11/21/2018Terminated
Z-0668-2019 Class IIAlinity i Progesterone Reagent Kit, List Number 08P3620Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the… 11/21/2018Terminated
Z-0669-2019 Class IIAlinity i 25-OH Vitamin D Reagent Kit, List Number 08P4532Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the… 11/21/2018Terminated