Recall Z-0671-2008

Parks Medical Electronics Inc · initiated November 8, 2007

Terminated

Product

Parks Medical Ultrasonic Doppler Flow Detector with Cautery Suppressor, Model 811-BL (finished assembly without accompanying probes), Parks Medical Electronics, Inc., Aloha, OR 97007

Codes / lots: Model 811-BL. Serial Number: 811-B-400001438

Reason for recall

Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.

FDA-determined cause: Process control

Action taken

811-B: On 9/28/07 Parks Medical Electronics notified their related U.S. distributor, Parks Medical Electronics Sales, Inc., Las Vegas, NV by email instructing them to stop further shipments. On 10/1/07 firm followed up with phone call instructing all units be returned for rework. On 11/8/07 firm sent replacement units accompanied with a Warranty Exchange Letter to their other U.S. customers notifying them of affected units. Letters were customized for each customer with a fill-in-the-blank of the model and serial numbers. Letters and replacement units were sent UPS 2nd Day Air and informed customer about the assembly error and instructions for returning the defective units. On 11/8/07 emails were sent to Brazilian and Australian distributors informing about assembly error and affected units. Rework instructions and a schematic were provided by email for each distributor to perform their own rework. On 11/28/97 Canadian customers were sent Warranty Exchange Letters and replacement units via UPS Ground. Firm is retrieving units from US and Canadian firms and will rework them. Firm is waiting to hear whether Brazilian or Austraian distributors need to sub-recall. Canadian distributors instructed to conduct subrecall if needed. All returned units will be reworked by disassembly of unit, removal of affected printed circuit board, washing board to remove acid flux, and testing.

Details

Recall number
Z-0671-2008
Initiated
November 8, 2007
Posted by FDA
July 3, 2008
Terminated
October 13, 2011
Product code
DPW – Flowmeter, Blood, Cardiovascular
Quantity
1 unit distributed to: Canadian veterinarian hospital on 9/26/07
Distribution
Worldwide Distribution - USA, Brazil, Australia, and Canada
Recalling firm
Parks Medical Electronics Inc, Aloha, OR

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0670-2008 Parks Medical Ultrasonic Doppler Flow Detector, Model 811-B (finished assembly without…Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure… 11/08/2007Terminated
Z-0672-2008 Parks Medical Ultrasonic Doppler Flow Detector, Model 811-BTS (finished assembly without…Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure… 11/08/2007Terminated
Z-0673-2008 Parks Medical Dual Frequency Doppler Vascular Flow Detector, Model 915-BL (finished…Premature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure… 11/08/2007Terminated