Recall Z-0678-2017

Teleflex Medical · initiated November 8, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

OnControl Biopsy System Ported Needle Tray, Sterile, Rx only, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marrow aspiration of the iliac crest of adult and pediatric patients. The Arrow OnControl Bone Marrow Biopsy System is intended for bone marrow core biopsy of the anterior or posterior iliac crest of adult patients

Codes / lots: Material # 9470-VC-006, Lot numbers: 010859 & 011511.

Reason for recall

Sterility: Due to a potential incomplete seal on the outer sterile package.

FDA-determined cause: Device Design

Action taken

Teleflex sent an Urgent Medical Device Recall Notification letter dated November 8, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product, complete the enclosed Recall Acknowledgement Form and fax to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call their local sales representative or Customer Service at 1-866-246-6990. For questions regarding this recall call 919-433-4940.

Details

Recall number
Z-0678-2017
Initiated
November 8, 2016
Terminated
November 20, 2019
Product code
KNW – Instrument, Biopsy
Distribution
Worldwide Distribution - US (nationwide) Internationally to Brazil, Canada, Ecuador, Guatemala, Mexico and Peru.
Recalling firm
Teleflex Medical, Research Triangle Park, NC

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0675-2017 Class IIARROW OnControl Bone Marrow Biopsy System Tray, Sterile, Rx only, Manufactured for:…Sterility: Due to a potential incomplete seal on the outer sterile package. 11/08/2016Terminated
Z-0676-2017 Class IIARROW OnControl Bone Marrow Biopsy System Comprehensive Tray, Sterile, The Arrow…Sterility: Due to a potential incomplete seal on the outer sterile package. 11/08/2016Terminated
Z-0677-2017 Class IIARROW OnControl Bone Lesion Biopsy System Tray, Sterile, Rx only, The Arrow OnControl…Sterility: Due to a potential incomplete seal on the outer sterile package. 11/08/2016Terminated
Z-0679-2017 Class IIARROW OnControl Ported Aspiration System Tray, Sterile, Rx only, The Arrow OnControl…Sterility: Due to a potential incomplete seal on the outer sterile package. 11/08/2016Terminated