Recall Z-0682-2024
Abbott Laboratories · initiated October 19, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ARCHITECT c4000 Processing Module REF: 02P24-01 02P24-02 02P24-40 01R24-56 01R25-56 The Abbott ARCHITECT c4000 Processing Module is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology.
Codes / lots: List Number: 02P24-01 Serial Numbers / UDI Codes: C400911 (01)00380740003746(21)C400911 C402913 (01)00380740003746(21)C402913 C460004 (01)00380740003746(21)C460004 C460033 (01)00380740003746(21)C460033 C460055 (01)00380740003746(21)C460055 C460066 (01)00380740003746(21)C460066 C460082 (01)00380740003746(21)C460082 C460158 (01)00380740003746(21)C460158 C460200 (01)00380740003746(21)C460200 C460375 (01)00380740003746(21)C460375 C460385 (01)00380740003746(21)C460385 C460396 (01)00380740003746(21)C460396 C460397 (01)00380740003746(21)C460397 C460408 (01)00380740003746(21)C460408 C460410 (01)00380740003746(21)C460410 C460422 (01)00380740003746(21)C460422 C460435 (01)00380740003746(21)C460435 C460439 (01)00380740003746(21)C460439 C460523 (01)00380740003746(21)C460523 C460532 (01)00380740003746(21)C460532 C460588 (01)00380740003746(21)C460588 C460728 (01)00380740003746(21)C460728 C460767 (01)00380740003746(21)C460767 C460797 (01)00380740003746(21)C460797 C460798 (01)00380740003746(21)C460798 C460800 (01)00380740003746(21)C460800 C460826 (01)00380740003746(21)C460826 C460827 (01)00380740003746(21)C460827 C460830 (01)00380740003746(21)C460830 C460832 (01)00380740003746(21)C460832 C460833…
Reason for recall
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
FDA-determined cause: Under Investigation by firm
Action taken
On 10/19/2023, the firm sent an "Urgent Medical Device Correction" Letter via FedEx priority overnight express to customers informing them that Abbott has determined that, A labeling update is required for the products listed above to meet US regulatory requirement 21 CFR 801.437 (User labeling for devices that contain natural rubber) and Brazilian requirement RDC 37/2015 (medical devices and IVDs to include in the label a Standardized Latex Warning in Portuguese CONTAIN NATURAL LATEX. MAY CAUSE ALLERGIES ). These Abbott products include internal hardware parts and/or subassemblies that contain dry natural rubber (latex) which may cause allergic reactions, but do not include the required precautionary label. Additionally, the Alinity s System Solid Waste Container Rubber Band (LN 04U97-01), used to secure the biohazard bag to the solid waste container, contains dry natural rubber (latex). Customer are instructed to: 1) Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the above listed Abbott product(s). 2) If you have forwarded the product(s) listed above to other laboratories, please inform them of this Product Correction and provide a copy of this letter. 3) Please retain this letter for your laboratory records. A precautionary label will be applied to the affected products. An Abbott representative will schedule a service visit when the instrument label is available for application. For questions or concerns, contact Customer Service at 1-877-4ABBOTT (24 hr/7 days a week).
Details
- Recall number
- Z-0682-2024
- Initiated
- October 19, 2023
- Posted by FDA
- January 9, 2024
- Product code
- JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Quantity
- 995 systems
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV O.U.S.: Brazil
- Recalling firm
- Abbott Laboratories, Irving, TX
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0680-2024 Class II | Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module is a fully…The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry… | 10/19/2023Open, Classified |
| Z-0681-2024 Class II | Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully…The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry… | 10/19/2023Open, Classified |