Recall Z-0684-2012

ConMed Corporation · initiated April 15, 2011

Terminated

Product

STAT 2 Pumpette Extension Set Gravity Flow Compensating Controller (Needle-Free), 12 length, Catalog/REF No. P-S2-12 N, STERILE, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. Product Usage: The STAT 2 Pumpette Extension sets are for use in conjunction with commonly available I.V. administration sets for controlling the infusion of I.V. fluids into the human body. The STAT 2 Pumpette Primary Administration Sets are for use in attachment to an I.V. bag and patient catheter inserted into a vein to control the infusion of intravenous fluids into the human body. This is a disposable device with a sterile, nonpyrogenic fluid path.

Codes / lots: Beginning Lot Code 0603241, Ending Lot Code 1103114

Reason for recall

ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the body of the compensating controller. The investigation of the complaints determined that the STAT 2 Pumpette Compensating Controller on several complaint samples had been assembled incorrectly.

FDA-determined cause: Device Design

Action taken

ConMed sent an "Urgent Medical Device Recall" letters and Reply Forms dated April 15, 2011 to the domestic consignees via UPS Priority Mail pouches and to the foreign consignees via FedEx International Priority. The letter identified the product, description of problem, and actions to be taken. Consignees were instructed to review their inventory and immediately stop use of the affected devices and return them to ConMed Corporation. Consignees who had distributed the affected products outside of their facility were instructed to immediately notify their customers of the recall and provide them with a copy of this notice. Consignees were asked to complete and return the enclosed Reply Form. For questions call 315-624-3237 or email ivcontroller@conmed.com

Details

Recall number
Z-0684-2012
Initiated
April 15, 2011
Posted by FDA
January 11, 2012
Terminated
November 20, 2012
Product code
FPA – Set, Administration, Intravascular
Quantity
1,146,741 sets in total
Distribution
Worldwide Distribution - USA (nationwide) - including the countries of: Australia, Barbados, Belgium, Bermuda, Canada, Egypt, Israel, India, Italy, Korea, New Zealand, Philippines and Saudi Arabia.
Recalling firm
ConMed Corporation, Utica, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0682-2012 STAT 2 Pumpette I.V. Compensating Controller, Catalog/REF No. PS2, ASSEMBLED IN MEXICO…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated
Z-0683-2012 STAT 2 Pumpette Extension Set Gravity Flow Compensating Controller, 12" length…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated
Z-0685-2012 STAT 2 Pumpette Primary Administration Set Gravity Flow Compensating Controller, 20…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated
Z-0686-2012 STAT 2 Pumpette Primary Administration Set Gravity Flow Compensating Controller…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated
Z-0687-2012 STAT 2 Pumpette Primary Administration Set Gravity Flow Compensating Controller, 60…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated
Z-0688-2012 STAT 2 Pumpette Primary Administration Set Gravity Flow Compensating Controller…ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the… 04/15/2011Terminated