Recall Z-0685-2026

B Braun Medical Inc · initiated October 29, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMN SET W/15 DROP, ULTRAPORT,134 IN.; Catalog Number: 490312. 2. IV ADMIN SET W/ 15 DROPS/ML, 120 IN.; Catalog Number: 490432.

Codes / lots: 1. Catalog Number: 490312; Primary UDI-DI: 04046964525643; Unit of Dose UDI-DI: 04046964525636; Expiration Date: Earliest Exp of Comp or 36mths. 2. Catalog Number: 490432; Primary UDI-DI: 04046964832826; Unit of Dose UDI-DI: 04046964832819; Expiration Date: Earliest Exp of Comp or 36mths.

Reason for recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

FDA-determined cause: Process control

Action taken

B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."

Details

Recall number
Z-0685-2026
Initiated
October 29, 2025
Posted by FDA
November 26, 2025
Product code
FPA – Set, Administration, Intravascular
Quantity
226,632 units
Distribution
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Recalling firm
B Braun Medical Inc, Bethlehem, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0603-2026 Class IIADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0604-2026 Class IIADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0605-2026 Class IIADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0606-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0607-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0608-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0609-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0610-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0611-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0612-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0613-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0614-2026 Class IIAnesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0615-2026 Class IIB.BRAUN utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0616-2026 Class IIBlood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0617-2026 Class IIBlood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0618-2026 Class IIBlood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0619-2026 Class IIBlood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0620-2026 Class IIBlood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0621-2026 Class IIBurette Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0622-2026 Class IIBurette Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0623-2026 Class IICARESAFE utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0624-2026 Class IICaresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0625-2026 Class IICaresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0626-2026 Class IICaresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0627-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0628-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0629-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0630-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0631-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified
Z-0632-2026 Class IIExtension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… 10/29/2025Open, Classified