Recall Z-0686-2020
Polymer Technology Systems · initiated December 2, 2016
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CVS Health At Home A1C Test Kit, Software Version Revision D, REF Number 3059
Codes / lots: Lot Numbers: 1606319 1606405 1606460 1610622 1610910 1611163 1618153
Reason for recall
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
FDA-determined cause: Software design
Action taken
A Customer Bulletin dated November 2016 was distributed to customers.
Details
- Recall number
- Z-0686-2020
- Initiated
- December 2, 2016
- Terminated
- July 16, 2020
- Product code
- LCP – Assay, Glycosylated Hemoglobin
- Quantity
- 8285
- Distribution
- The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya…
- Recalling firm
- Polymer Technology Systems, Whitestown, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0682-2020 Class II | A1CNow+ Systems (professional use product), Software Version Revision D, REF Numbers US:…A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13%… | 12/02/2016Terminated |
| Z-0683-2020 Class II | A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13%… | 12/02/2016Terminated |
| Z-0684-2020 Class II | Walgreens At-Home A1C Test Kit, Software Version Revision D, REF Number 3056A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13%… | 12/02/2016Terminated |
| Z-0685-2020 Class II | ReliOn FastA1C Test, Software Version Revision D, REF Number 3055A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13%… | 12/02/2016Terminated |