Recall Z-0691-2026
B Braun Medical Inc · initiated October 29, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2 CARESITES, 125 IN.; Catalog Number: 490408.
Codes / lots: 1. Catalog Number: 490408; Primary UDI-DI: 04046964773266; Unit of Dose UDI-DI: 04046964773259; Expiration Date: Earliest Exp of Comp or 24mths.
Reason for recall
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA-determined cause: Process control
Action taken
B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."
Details
- Recall number
- Z-0691-2026
- Initiated
- October 29, 2025
- Posted by FDA
- November 26, 2025
- Product code
- FPA – Set, Administration, Intravascular
- Quantity
- 24,072 units
- Distribution
- Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
- Recalling firm
- B Braun Medical Inc, Bethlehem, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0603-2026 Class II | ADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0604-2026 Class II | ADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0605-2026 Class II | ADDitIV utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0606-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0607-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0608-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0609-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0610-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0611-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0612-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0613-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0614-2026 Class II | Anesthesia IV Set utilized in gravity IV administration sets and pump administration…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0615-2026 Class II | B.BRAUN utilized in gravity IV administration sets and pump administration sets utilized…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0616-2026 Class II | Blood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0617-2026 Class II | Blood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0618-2026 Class II | Blood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0619-2026 Class II | Blood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0620-2026 Class II | Blood Administration Set utilized in gravity IV administration sets and pump…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0621-2026 Class II | Burette Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0622-2026 Class II | Burette Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0623-2026 Class II | CARESAFE utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0624-2026 Class II | Caresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0625-2026 Class II | Caresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0626-2026 Class II | Caresite utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0627-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0628-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0629-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0630-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0631-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |
| Z-0632-2026 Class II | Extension Set utilized in gravity IV administration sets and pump administration sets…Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and… | 10/29/2025Open, Classified |