Recall Z-0694-2015
Stryker Endoscopy · initiated November 17, 2014
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080360. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Codes / lots: 614661 716054 817474 913490 913546 913557 913558 913564 913652 914124 915684 915912 915913 1013456 1013457 1014295 1014657 1014658 1015047 1015048 1015158 1015160 1015781 1015782 1016402 1016403 1016404 1016557 1016558 1016559 1016561 1016828 1016829 1113336 1113337 1113338 1113615 1113616 1113698 1113699 1114268 1114269 1115399 1115401 1115402 1115964 1115965 1115966 1116116 1116117 1116118 1117009 1117010 1213733 1213734 1214732 1214733 1216815 1216816 1216817 1216818 1217032 1313934 1314205 1314921 1314922 1314923 1315783 1315784 1316806 1316807 1410093 1410094 1410095 1410417 1410418 1410599 1410600 1410601 1412348
Reason for recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
FDA-determined cause: Error in labeling
Action taken
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Details
- Recall number
- Z-0694-2015
- Initiated
- November 17, 2014
- Posted by FDA
- December 12, 2014
- Terminated
- July 17, 2015
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South…
- Recalling firm
- Stryker Endoscopy, San Jose, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0592-2015 Class II | PKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide:…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0593-2015 Class II | PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080120 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0594-2015 Class II | PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0595-2015 Class II | PKG, BIPOLAR FORCEPS, MICRO TIP, 33CM, P/N 0250080122 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0596-2015 Class II | PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0597-2015 Class II | PKG, OP-BIOPSY FORCEPS, SPOON JAW, P/N 0250080007 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0598-2015 Class II | PKG, FIXATION CLAMP 5MM WITH RATCHET, P/N 0250080130 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0599-2015 Class II | PKG, KNOT PUSHER 5MM OPEN, P/N 0250080132 Laparoscopic Manual Instruments are intended…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0600-2015 Class II | PKG, MYOMA DRILL 5MM LENGTH 320MM, P/N 0250080133 Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0601-2015 Class II | PKG, PALPATION PROBE GRADUATED 5MM, P/N 0250080135; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0602-2015 Class II | PKG, PORTIO ACORN SMALL ONLY, P/N 0250080139; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0603-2015 Class II | PKG, TROCAR 5.5MM PYRAMIDAL TIP, P/N 0250080154; Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0604-2015 Class II | PKG, OBTURATOR, 5.5MM FOR HASSON CANNULA, P/N 0250080157; Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0605-2015 Class II | PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0606-2015 Class II | PKG, MARYLAND FORCEPS, CURVED, P/N 0250080009. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0607-2015 Class II | PKG, HASSON CANNULA, 5.5MM, AUTOMATIC VALVE, STOP COCK, P/N 0250080176. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0608-2015 Class II | PKG, MIXTER FORCEPS, P/N 0250080010. Laparoscopic Manual Instruments are intended for…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0609-2015 Class II | PKG, HOOK SCISSORS, P/N 0250080020. Laparoscopic Manual Instruments are intended for…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0610-2015 Class II | PKG, HASSON CANNULA, 11MM AUTOMATIC VALVE, STOP COCK, P/N 0250080179. Laparoscopic…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0611-2015 Class II | PKG, FIXABLE CONE FOR 11MM HASSON, P/N 0250080180. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0612-2015 Class II | PKG, METZENBAUM SCISSORS, STRAIGHT, P/N 0250080024. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0613-2015 Class II | PKG, METZENBAUM SCISSORS, CURVED, LONG, P/N 0250080025. Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0614-2015 Class II | PKG, BABY METZENBAUM SCISSORS, CURVED, P/N 0250080028. Laparoscopic Manual Instruments…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0615-2015 Class II | PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080029. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0616-2015 Class II | PKG, 12.5/11MM - 5.5 REDUCER, P/N 0250080204. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0617-2015 Class II | PKG, BULLET FORCEPS, TAPERED, SERRATED, SPOON, P/N 0250080041. Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0618-2015 Class II | PKG, LUERLOCK, MALE, 6MM TUBING, P/N 0250080207. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0619-2015 Class II | PKG, LUERLOCK, FEMALE, 6MM TUBING, P/N 0250080208. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0620-2015 Class II | PKG, MARYLAND FORCEPS, CURVED, LEFT, P/N 0250080042. Laparoscopic Manual Instruments are…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |
| Z-0621-2015 Class II | PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080230. Laparoscopic Manual…The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070… | 11/17/2014Terminated |