Recall Z-0696-2013

Roche Diagnostics Operations, Inc. · initiated November 19, 2012

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.

Codes / lots: Part Numbers: 28065047001 and 03245233001

Reason for recall

A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access

FDA-determined cause: Software design

Action taken

Roche sent an Urgent Medical Device Correction letter on November 19, 2012, to all affected customers via UPS Ground (receipt required). The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the attached form and return by fax and file this Medical Device Correction for future reference. For technical support customers were instructed to contact Diagnostics Technical Support at 1-900-428-2336. For questions regarding this recall call 317-521-3911.

Details

Recall number
Z-0696-2013
Initiated
November 19, 2012
Posted by FDA
January 18, 2013
Terminated
January 13, 2014
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
845 Units
Distribution
Nationwide Distribution including AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR.
Recalling firm
Roche Diagnostics Operations, Inc., Indianapolis, IN