Recall Z-0696-2024

Medtronic Xomed, Inc. · initiated November 7, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NIM TriVantage EMG Endotracheal Tubes, Model (REF) Numbers and sizes: a) 8229706, 6.0MM ID; b) 8229707, 7.0MM ID; c) 8229708, 8.0MM ID; d) 8229736, 6MM ID 3PK; e) 8229737, 7MM ID 3 PK; f) 8229738, 8MM ID 3 PK

Codes / lots: a) 8229706, GTIN 00643169782587, Serial Numbers: 0225892912, 0225894523, 0225904858, 0226227428, 0226246448, 0226246716, 0226253848, 0226258673, 0226276006, 0226294121, 0226299303, 0226299304, 0226321727, 0226350207, 0226604403, 0226628515, 0226648757, 0226684520, 0226718176, 0226745452; GTIN 00763000033941, Serial Numbers: 0225750658, 0225785080, 0225785081, 0225785083, 0225785085, 0225785587, 0225787473, 0225974682, 0225998102, 0225998104, 0226099513, 0226121951, 0226130941; b) 8229707, GTIN 00643169373099, Serial Numbers: 0225784379, 0225784380, 0225798573, 0225809609, 0225832077, 0225841439, 0225848733, 0225877835, 0225952300, 0225952301, 0225994034, 0226020502, 0226030100, 0226121428, 0226124551, 0226130051, 0226130055, 0226150740, 0226166004, 0226337996; GTIN 00643169782594, Serial Numbers: 0225798573, 0225803180, 0225805468, 0225805470, 0225824531, 0225825616, 0225828951, 0225832074, 0225832077, 0225841440, 0225846915, 0225846916, 0225848733, 0225848736, 0225848738, 0225849141, 0225869998, 0225870000, 0225877834, 0225877835, 0225886691, 0225915233, 0225915542, 0225921185, 0225929260, 0225929261, 0225932768, 0225934163, 0225952300, 0225952301, 0225956683, 0225962826…

Reason for recall

Firm received reports of customers experiencing noise from the NIM System, lead-off or high impedance issues, or loss of/intermittent nerve monitoring,

FDA-determined cause: Under Investigation by firm

Action taken

Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/07/2023 via 2-day UPS mailing. The notice explained the problem, the potential health risk, and requested segregation and quarantine pending return of the affected product. For questions, contact: neuro-quality@medtronic.com.

Details

Recall number
Z-0696-2024
Initiated
November 7, 2023
Posted by FDA
January 10, 2024
Product code
ETN – Stimulator, Nerve
Quantity
80569 units
Distribution
Worldwide distribution.
Recalling firm
Medtronic Xomed, Inc., Jacksonville, FL

Official FDA recall record