Recall Z-0698-2007

Boston Scientific Corporation · initiated February 7, 2007

Terminated

Product

Boston Scientific SpyGlass Light Cable Reference Number: 4621 UPN M00546210

Codes / lots: Lot Number: 355R

Reason for recall

Light Degradation in transmission-A film may deposit on the front face and limit transmission of light

FDA-determined cause: Other

Action taken

Boston Scientific notified accounts by letter dated 2/1/07 requesting return of product.

Details

Recall number
Z-0698-2007
Initiated
February 7, 2007
Posted by FDA
March 31, 2007
Terminated
November 15, 2007
Product code
EQH – Source, Carrier, Fiberoptic Light
Quantity
13 units
Distribution
Nationwide Distribution --- including states of CA, CO, FL, OH, MA, MN, MO NY, PA, SC,TX, and WA .
Recalling firm
Boston Scientific Corporation, Natick, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0696-2007 Boston Scientific SpyGlass Visualization Probe Reference Number: 4603 UPN M00546030SpyGlass Probe and Ocular: Ocular is out of specification and during disconnect may damage probe 02/07/2007Terminated
Z-0697-2007 Boston Scientific SpyGlass Ocular Reference Number: 4604 UPN M00546040SpyGlass Probe and Ocular: Ocular is out of specification and during disconnect may damage probe 02/07/2007Terminated
Z-0699-2007 Boston Scientific SpyGlass Camera Ref Number: 4610 UPN M00546100A teal "bloom"/ glare on the subject image during high light conditions. 02/07/2007Terminated