Recall Z-0703-2022
Microbiologics Inc · initiated December 1, 2015
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture). Sets include: 1. 0864E3 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356
Codes / lots: Catalog Number: 8164; Lot Number (Expiration Date): 8164-45 (12/16/2015), 8164-46 (12/16/2015), 8164-47 (09/02/2016), 8164-48 (09/02/2016), and 8164-49 (10/03/2016); UDI: 70845357028623
Reason for recall
The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.
FDA-determined cause: Material/Component Contamination
Action taken
Microbiologics notified customers on about 12/01/2015, via letter titled "MEDICAL DEVICE RECALL." Instructions to customers included verify lots in inventory, cease use of affected product, quarantine affected product, arrange for the return of affected product, and complete and return the response form.
Details
- Recall number
- Z-0703-2022
- Initiated
- December 1, 2015
- Terminated
- July 7, 2022
- Product code
- JTR – Kit, Quality Control For Culture Media
- Quantity
- 541 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MN, NC, NM, NY, OH, OR, PA, TN, TX, VA, VT, WA, WI, WV, WY.
- Recalling firm
- Microbiologics Inc, Saint Cloud, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0704-2022 Class II | QC Sets and Panels: KWIK-STIK Cepheid Smart GBS QC Set (Live Culture). Sets include: 1.…The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid… | 12/01/2015Terminated |