Recall Z-0703-2024
BioFire Diagnostics, LLC · initiated November 27, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems.
Codes / lots: UDI-DI: 00815381020109. Kit Lot/Pouch Lot/Expiration: 1749723/2XKB23/July 6, 2024; 1746223/2XJA23/July 6, 2024
Reason for recall
A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.
FDA-determined cause: Process control
Action taken
On 11/27/23, recall notices were emailed to laboratory directors who were asked to do the following: Discontinue use and discard any remaining pouches that are subject to this recall. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. Complete and return the acknowledgement form using the following URL: https://www.biofiredx.com/recall5797 Customers with questions can contact Technical Support at biofiresupport@biomerieux.com or via telephone by dialing 1.800.735.6544 and selecting option 5.
Details
- Recall number
- Z-0703-2024
- Initiated
- November 27, 2023
- Posted by FDA
- January 10, 2024
- Product code
- PCH – Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
- Quantity
- 53 Kits
- Distribution
- US: NC, MO, NE, CT, IL, TN, NJ, OH, MA, CA, FL, UT, IA, KS, TX, MD
- Recalling firm
- BioFire Diagnostics, LLC, Salt Lake City, UT
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0704-2024 Class II | The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test…A manufacturing event may contribute to elevated rates of false negative results while using… | 11/27/2023Open, Classified |