Recall Z-0704-2025
GE OEC Medical Systems, Inc · initiated November 21, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Codes / lots: Product Name: OEC Elite GTIN code: 00001234567895 Lot/Serial Numbers / UDI codes: FAXXXE00254 (01)00840682121729(11)190400(21)FAXXXE00254 FBHXTX00837 (01)00840682121729(11)230300(21)FBHXTX00837 FBHXTX00912 (01)00840682121729(11)230500(21)FBHXTX00912 FBHXTX01278 (01)00840682121729(11)240800(21)FBHXTX01278 FBXXXE00731 (01)00840682121729(11)190500(21)FBXXXE00731 FBMHTX00470 (01)00840682121729(11)240900(21)FBMHTX00470 F2XXTX00801 (01)00840682121729(11)240800(21)F2XXTX00801 FBHXXX00171 (01)00840682121729(11)180600(21)FBHXXX00171 FBXXXE00715 (01)00840682121729(11)190500(21)FBXXXE00715 FBMHTX00444 (01)00840682121729(11)240600(21)FBMHTX00444 FBXXTE02974 (01)00840682121729(11)240800(21)FBXXTE02974 FBMHTX00466 (01)00840682121729(11)240900(21)FBMHTX00466 FAXXTE00672 (01)00840682121729(11)221200(21)FAXXTE00672 F9XXTX00276 (01)00840682121729(11)230900(21)F9XXTX00276 FAXXTE01160 (01)00840682121729(11)240800(21)FAXXTE01160 FBMHTX00454 (01)00840682121729(11)240800(21)FBMHTX00454 FBMHTX00455 (01)00840682121729(11)240800(21)FBMHTX00455 FBMHTX00465 (01)00840682121729(11)240900(21)FBMHTX00465 FBXXTE02975 (01)00840682121729(11)240800(21)FBXXTE02975 FBXXTE02976…
Reason for recall
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
FDA-determined cause: Process control
Action taken
On 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative. 3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions. 4.Please complete and return the attached acknowledgement form torecall.15150@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed. for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Details
- Recall number
- Z-0704-2025
- Initiated
- November 21, 2024
- Posted by FDA
- December 18, 2024
- Product code
- OXO – Image-Intensified Fluoroscopic X-Ray System, Mobile
- Quantity
- 100 systems
- Distribution
- U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
- Recalling firm
- GE OEC Medical Systems, Inc, Salt Lake City, UT
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0703-2025 Class II | OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile…Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and… | 11/21/2024Open, Classified |
| Z-0705-2025 Class II | OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite…Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and… | 11/21/2024Open, Classified |
| Z-0706-2025 Class II | OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is…Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and… | 11/21/2024Open, Classified |