Recall Z-0714-2022

Microbiologics Inc · initiated October 6, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299

Codes / lots: Catalog Number HL163263 OEM Private Label with Healthlink; UDI: N/A (OEM agreement with Healthlink ended in Q1 2016 before UDI was required for class 1 IVDs) Lot Number (Expiration Date): 959-74-1 (10/08/2014), 959-75-2 (04/22/2015), 959-75-3 (04/22/2015), 959-75-6 (04/22/2015), 959-75-7 (04/22/2015), 959-75-10 (04/22/2015), and 959-76-2 (07/08/2015)

Reason for recall

This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.

FDA-determined cause: Nonconforming Material/Component

Action taken

Microbiologics notified customers on about 10/06/2014 via letter titled "E.faecalis Gentamicin Susceptibility Notification." Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall, it is recommended that the intended use of the organism be reviewed to determine if Gentamicin synergy results have an impact on use. It is recommended to discard these lots if an impact is to be expected. Questions or concerns can be directed to the Technical Support team at 320.229.7064 or techsupport@microbiologics.com.

Details

Recall number
Z-0714-2022
Initiated
October 6, 2014
Terminated
January 17, 2023
Product code
JTR – Kit, Quality Control For Culture Media
Quantity
150 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South…
Recalling firm
Microbiologics Inc, Saint Cloud, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0712-2022 Class IIKWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack…This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A… 10/06/2014Terminated
Z-0713-2022 Class IILYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus…This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A… 10/06/2014Terminated