Recall Z-0724-2017

Halyard Health, Inc · initiated November 9, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile), Product Code 44966 00; C-Section Drape with Clear Screen (Non-Sterile), Product Code 44967 NS; C-Section Drape with Clear Screen (Sterile), Product Code 44968 00; C-Section Drape with Pouch and Fenestration (Non-Sterile), Product Code 44977 NS; C-Section Drape with Pouch and Fenestration (Sterile), Product Code 44978 00 Product Usage: A surgical drape is a device made of natural or synthetic materials intended to be use as a protective patient covering, such as to isolate a site or surgical incision from microbial and other contamination.

Codes / lots: Product Code 44965 NS - Lot Numbers AC6174###, AC62266###, AC6229###, AC6243###, AC6252###; Product Code 44966 00 - Lot Numbers AC6146##L, AC6157##L, AC6177##L, AC6215##L, AC6221##L, AC6228##L, AC6245##L, AC6252##L, AC6258##L, AC6272##L, AC6277##L; Product Code 44967 NS - Lot Numbers AC6170###, AC6219###, AC6223###, AC6229###, AC625###; Product Code 44068 00 - Lot Numbers AC6146##L, AC6177##L, AC6230##L, AC6231##L, AC6234##L, AC6242##L, AC6245##L, AC6264##L, AC6271##L, AC6277##L; Product Code 44077 NS - Lot Numbers AC6170###, AC6213###, AC6214###, AC6222###, AC6223###, AC6251###; Product Code 44978 00 - Lot Numbers AC6147##L, AC6177##L, AC6243##L, AC 6252##L - (# = includes numbers 0, 1, 2, 3, 4, 5, 6, 7, 8 or 9 - L = includes letters a, b, or c)

Reason for recall

Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.

FDA-determined cause: Nonconforming Material/Component

Action taken

Accounts were notified of the recall by letter on 11/9/2016. The consignees were instructed to quarantine and discontinue use of the recalled lots and to discard/destroy per the facility's standard disposal procedure for nonwoven materials. A response form was included which was to be sent back.

Details

Recall number
Z-0724-2017
Initiated
November 9, 2016
Posted by FDA
December 8, 2016
Terminated
February 22, 2017
Product code
KKX – Drape, Surgical, Antimicrobial
Quantity
132 cases
Distribution
US Nationwide Distribution in the states of AL, FL, GA, IL, KS, MN, MO, NE, NJ, OH, PH, TN, VA, WI, WY
Recalling firm
Halyard Health, Inc, Alpharetta, GA

Official FDA recall record