Recall Z-0725-2026

Enterix, Inc. · initiated September 17, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.

Codes / lots: 1. Model Number: 90010; GTIN:10850093003313; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 2. Model Number: 90025; GTIN: 20850093003310; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 3. Model Number: 90030; GTIN: 00850093003279; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102.

Reason for recall

InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.

FDA-determined cause: Under Investigation by firm

Action taken

Consignees were initially sent an email as notification of the recall on about 09/17/2025, which was followed by a formal recall notification letter on about 09/23/2025. Consignees were instructed to immediately examine inventory and quarantine any affected units, notify customers if affected further distributed, and for those patients that have already been tested using an InSure ONE test kit that has been identified as part of the scope of this recall be retested. Once an inventory count has been received by Enterix, they will provide RMA and courier information so the recalled product can be returned for sorting and destruction.

Details

Recall number
Z-0725-2026
Initiated
September 17, 2025
Posted by FDA
November 28, 2025
Product code
KHE – Reagent, Occult Blood
Quantity
201,426 test kits
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.
Recalling firm
Enterix, Inc., Edison, NJ