Recall Z-0737-2018
Ventana Medical Systems Inc · initiated December 19, 2017
Class I – dangerous or defective products that could cause serious health problems or death
Product
OptiView Amplification Kit (250 Test), Ventana Part Number 860-099, Roche GMMI 06718663001 in vitro diagnostic
Codes / lots: UDI 04015630984749, Lot Numbers: Y19318
Reason for recall
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
FDA-determined cause: Process control
Action taken
On Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected product is available. Actions to be taken by the customer/user Affected kits may continue to be used by customers until corrected product is available, however Ventana is mandating that the affected IHC detection kits (iView, ultraView, OptiView) must only be used in conjunction with same-slide controls. These controls must be appropriate for each assay and capable of detecting false negative results due to a complete or partial reagent dispense failure. CINtec PLUS Cytology does not have the capacity for same slide controls, so system-level controls should be maintained. ultraView SISH Detection is used for HER2 analysis, and employs internal positive controls; no external control is required. For assays that directly relate to clinical therapy decision making (e.g. ER/PR, HER2, ALK, etc.), it is additionally important to select a same slide positive control tissue with sufficient sensitivity to detect small decreases in intensity that may cause borderline positive cases to appear as negative (e.g. HER2 2+ vs. 1+). Although the use of same slide controls is considered optimal laboratory practice and strongly recommended by Ventana, customers may revert to standard run controls once non-impacted product is received. In order to reduce the risk of this issue impacting patient care, customers not using same slide controls as a standard practice should follow their
Details
- Recall number
- Z-0737-2018
- Initiated
- December 19, 2017
- Posted by FDA
- February 27, 2018
- Terminated
- November 22, 2021
- Product code
- NJT – Immunohistochemistry Reagents And Kits
- Quantity
- 518 units
- Distribution
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
- Recalling firm
- Ventana Medical Systems Inc, Oro Valley, AZ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0729-2018 Class I | OptiView DAB IHC Detection Kit, Ventana Part Number 760-700, Roche GMMI 06396500001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0730-2018 Class I | ultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0731-2018 Class I | iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0732-2018 Class I | ultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0733-2018 Class I | ultraView SISH DNP Detection Kit US, Ventana Part Number 760-098, Roche GMMI 05572037001…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0734-2018 Class I | CINtec PLUS Cytology Kit (CE-IVD), Ventana Part Number 605-100, Roche GMMI 06889565001…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0735-2018 Class I | CINtec PLUS Cytology Kit (Canada/Japan), Ventana Part Number 805-100, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0736-2018 Class I | OptiView Amplification Kit, Ventana Part Number 760-099, Roche GMMI 06396518001 in vitro…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0738-2018 Class I | Hematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnosticIncreased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0739-2018 Class I | ultraView SISH DNP Detection Kit, Ventana Part Number 800-098, Roche GMMI 05907136001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |
| Z-0740-2018 Class I | NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… | 12/19/2017Terminated |