Recall Z-0737-2018

Ventana Medical Systems Inc · initiated December 19, 2017

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

OptiView Amplification Kit (250 Test), Ventana Part Number 860-099, Roche GMMI 06718663001 in vitro diagnostic

Codes / lots: UDI 04015630984749, Lot Numbers: Y19318

Reason for recall

Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.

FDA-determined cause: Process control

Action taken

On Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected product is available. Actions to be taken by the customer/user Affected kits may continue to be used by customers until corrected product is available, however Ventana is mandating that the affected IHC detection kits (iView, ultraView, OptiView) must only be used in conjunction with same-slide controls. These controls must be appropriate for each assay and capable of detecting false negative results due to a complete or partial reagent dispense failure. CINtec PLUS Cytology does not have the capacity for same slide controls, so system-level controls should be maintained. ultraView SISH Detection is used for HER2 analysis, and employs internal positive controls; no external control is required. For assays that directly relate to clinical therapy decision making (e.g. ER/PR, HER2, ALK, etc.), it is additionally important to select a same slide positive control tissue with sufficient sensitivity to detect small decreases in intensity that may cause borderline positive cases to appear as negative (e.g. HER2 2+ vs. 1+). Although the use of same slide controls is considered optimal laboratory practice and strongly recommended by Ventana, customers may revert to standard run controls once non-impacted product is received. In order to reduce the risk of this issue impacting patient care, customers not using same slide controls as a standard practice should follow their

Details

Recall number
Z-0737-2018
Initiated
December 19, 2017
Posted by FDA
February 27, 2018
Terminated
November 22, 2021
Product code
NJT – Immunohistochemistry Reagents And Kits
Quantity
518 units
Distribution
Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
Recalling firm
Ventana Medical Systems Inc, Oro Valley, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0729-2018 Class IOptiView DAB IHC Detection Kit, Ventana Part Number 760-700, Roche GMMI 06396500001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0730-2018 Class IultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0731-2018 Class IiView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0732-2018 Class IultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0733-2018 Class IultraView SISH DNP Detection Kit US, Ventana Part Number 760-098, Roche GMMI 05572037001…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0734-2018 Class ICINtec PLUS Cytology Kit (CE-IVD), Ventana Part Number 605-100, Roche GMMI 06889565001…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0735-2018 Class ICINtec PLUS Cytology Kit (Canada/Japan), Ventana Part Number 805-100, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0736-2018 Class IOptiView Amplification Kit, Ventana Part Number 760-099, Roche GMMI 06396518001 in vitro…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0738-2018 Class IHematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnosticIncreased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0739-2018 Class IultraView SISH DNP Detection Kit, Ventana Part Number 800-098, Roche GMMI 05907136001 in…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated
Z-0740-2018 Class INEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI…Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers… 12/19/2017Terminated