Recall Z-0738-2016
Horiba Instruments Inc · initiated January 11, 2016
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinders reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.
Codes / lots: All Lots
Reason for recall
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
FDA-determined cause: Under Investigation by firm
Action taken
Customer notification letters were sent on 01/11/16. Customers were provided the following instructions/actions: When performing tests that use the Tinder reaction, the blood draw should be done before treatment is given to the patient. Reagent inserts for the concerned HORIBA Medical Products have been updated and now include the following additional warning: Patients treated with N-Acetylcysteine (NAC) for Paracetamol overdose may generate a false low result. The letter request customers share the information with laboratory staff and retain the notification as part of their laboratory documentation. The letter states it is mandatory for customers to complete and return the enclosed response form within 10 days in order for the firm to verify the understanding of the notice. If any questions contact the local HORIBA Medical representative.
Details
- Recall number
- Z-0738-2016
- Initiated
- January 11, 2016
- Posted by FDA
- February 2, 2016
- Terminated
- June 7, 2016
- Product code
- CHH – Enzymatic Esterase--Oxidase, Cholesterol
- Quantity
- 216
- Distribution
- U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
- Recalling firm
- Horiba Instruments Inc, Irvine, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0739-2016 Class II | ABX PENTRA Enzymatic Creatinine CP ref. A11A01907 reagent is intended for the…N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with… | 01/11/2016Terminated |
| Z-0740-2016 Class II | ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in…N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with… | 01/11/2016Terminated |
| Z-0741-2016 Class II | ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in…N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with… | 01/11/2016Terminated |
| Z-0742-2016 Class II | ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the…N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with… | 01/11/2016Terminated |
| Z-0743-2016 Class II | ABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the…N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with… | 01/11/2016Terminated |