Recall Z-0739-2020

Ortho Clinical Diagnostics Inc · initiated October 14, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

7% BSA (Bovine Serum Albumin), Product Code: 8262487 - Product Usage: is used to dilute samples when assay values exceed the reportable (dynamic) range using the VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Codes / lots: Lot with Expiry Date: D6289 2019-11-20 E6382 2020-01-15 F6602 2020-05-02 G6679 2020-08-13 H6897 2020-10-01 J7129 2021-01-23 K7421 2021-07-01 L7592 2021-08-26

Reason for recall

Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7 days

FDA-determined cause: Device Design

Action taken

On August 9, 2019 the firm distributed Urgent Medical Device Recall Communication letters by via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) summarizing the reason as Ortho Clinical Diagnostics determined that VITROS 7% BSA and VITROS Urine Electrolyte Diluent (UED) do not meet the current OAS time of d 7 days when stored on the VITROS 250/350 System in a 2mL cup only. Orthos internal testing compared results from samples diluted with diluent stored on the analyzer versus fresh diluent. When using 7% BSA or UED stored in a 2mL cup for 7 days, the differences in the results vs fresh diluent were >10% for some assays. It explained as a resolution: " The OAS of 7% BSA and UED in 2mL cups has been revised to d24 hours. " Current OAS time remains at d7 days for storage in 5mL and 10 mL bottles. " The Instructions for Use and appropriate User Documentation will be revised at a future date. It instructs them to: - Send the enclosed customer letter to all customers to whom you shipped or know to have a VITROS 250/350 System. Please continue to provide this letter to new customers until the Instructions for Use are updated. - Complete and return the Confirmation of Receipt form by October 22, 2019.

Details

Recall number
Z-0739-2020
Initiated
October 14, 2019
Terminated
November 13, 2020
Product code
DCF – Albumin, Antigen, Antiserum, Control
Quantity
16,584
Distribution
Worldwide distribution - US Nationwide distribution and countries BM, AU, BR, CA, CL, CN, CO, IN, JP, MX, SG, GB, FR, DE, IT, ES, PT, PL, RU, DK, NO, SE, BE, NL.
Recalling firm
Ortho Clinical Diagnostics Inc, Raritan, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0740-2020 Class IIUrine Electrolyte Diluent, Product Code: 1112352 - Product Usage: used to dilute urine…Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine… 10/14/2019Terminated