Recall Z-0740-2014

Atrium Medical Corporation · initiated November 19, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with in-line connectors Product Code: 3650-100. To evacuate air and/or fluid from the chest cavity or mediastinum.

Codes / lots: Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)

Reason for recall

Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

FDA-determined cause: Device Design

Action taken

Atrium Medical issued notification letters to direct consignees by UPS next day on 19 November 2013, and indirect consignees (distributors customers) by UPS next day or USPS priority (for customers with a PO Box) on 27 November 2013. The Letter explain the issue and the steps that should be taken. The customer will be offered replacement with unaffected product or credit. The firm describes Procedures for Emergency Use. The firm identifies potential product substitutions. These products are the same drainage product without the auto-transfusion line. International Distributors. International distributors provided with the Recall Authorization' package and the customer letter. Accounts are requested to complete the Recall Reply Form acknowledging receipt of the notification. Contact Atrium Medical Customer Service at 1-800-370-7899, Monday through Friday between 9:00 am to 5:00 pm.

Details

Recall number
Z-0740-2014
Initiated
November 19, 2013
Posted by FDA
January 14, 2014
Terminated
June 11, 2015
Product code
CAC – Apparatus, Autotransfusion
Quantity
42,699 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.
Recalling firm
Atrium Medical Corporation, Hudson, NH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0735-2014 Class IIAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0736-2014 Class IIAtrium Ocean Water Seal Chest Drain ATS Blood Recovery System with stopcock and in-line…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0737-2014 Class IIAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0738-2014 Class IIAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0739-2014 Class IIAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0741-2014 Class IIAtrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0742-2014 Class IIAtrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain; Product…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated
Z-0743-2014 Class IIAtrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with…Chest Drain tubing of of the ATS Blood recovery may leak or disconnect 11/19/2013Terminated