Recall Z-0756-2025

MEDLINE INDUSTRIES, LP - Northfield · initiated November 12, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ST OR MAYO HEGAR NEEDLEHOLDER 6"(OR2500), Medline Item No. 67870

Codes / lots: UDI: 00653160196894 (ea) 30653160196895 (case); Lot 2023091290

Reason for recall

Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to increased risk of contamination and infection.

FDA-determined cause: Process control

Action taken

On November 25, 2024, the firm notified customers via letters titled IMMEDIATE ACTION REQUIRED. Customers were informed that specific items and lots of Pure Pouch Single-Sterile and Sterile Kits are being recalled due to weak seals. Affected product must be destroyed. The letter included instructions on actions to take to be issued a credit. Customers should immediately check their stock for affected product and quarantine all units. Affected lots should be destroyed. When the recalling firm has received your completed destruction form, your account will be issued a credit, if applicable. If you have any questions, please contact 866-359-1704.

Details

Recall number
Z-0756-2025
Initiated
November 12, 2024
Posted by FDA
December 23, 2024
Product code
GEN – Forceps, General & Plastic Surgery
Quantity
21,440 total
Distribution
US Nationwide distribution.
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0754-2025 Class IISTERILE STRT KELLY HEMOSTAT (SK8039S), Medline Item No. 66695Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified
Z-0755-2025 Class IIO.R. SCISSORS, S/B, 5.5 STERILE, Medline Item No. DYND04000Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified
Z-0757-2025 Class IISTERILE MAGILL FORCEP (MAGF9), Medline Item No. 66790Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified
Z-0758-2025 Class IIST REPROCESSED BRUNS CURETTE #00 (BC00), Medline Item No. 67600KWeak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified
Z-0759-2025 Class IIST IUD REMOVAL HOOK 10" (504090), Medline Item No. I68785Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified
Z-0760-2025 Class IIRUMMELL, Medline Item No. ST006STWeak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to… 11/12/2024Open, Classified