Recall Z-0757-2013

Abbott Vascular · initiated November 30, 2011

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55, 1011650-65, 1011651-45, 1011651-55. The RX Accunet EPS is indicated for use as a guide wire and embolic protection system.

Codes / lots: Lot Number: 1081061, 1082561, 1091361, 1080561, 1081761, 1090261, 1091561, 1100461, 1080361, 1081661, 1090161, 1090861, 1092361, 1080861, 1082661, 1100761, 1081761, 1092861, 1082461, 1081161, 1082261, 1083061, 1081961, 1093061.

Reason for recall

The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall thickness.

FDA-determined cause: Nonconforming Material/Component

Action taken

Abbott Vascular sent an "URGENT DEVICE RECALL" letter dated November 30, 2011 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to work with their local account representative to review their inventory, complete the attached Field Action Reconciliation/ Effectiveness Check Form and return any unused identified products to Abbott Vascular.

Details

Recall number
Z-0757-2013
Initiated
November 30, 2011
Posted by FDA
January 31, 2013
Terminated
January 31, 2013
Product code
NTE – Temporary Carotid Catheter For Embolic Capture
Quantity
1253 units (962 units in the US & Puerto Rico)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, China, Germany, Spain, France, United Kingdom, Italy, Pakistan, Poland, Russia, Slovakia, Taiwan, and South Africa.
Recalling firm
Abbott Vascular, Temecula, CA

Official FDA recall record