Recall Z-0765-2017

Stryker Sustainability Solutions · initiated November 11, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Reprocessed Xcel Bladeless Trocar (H12LP) Model # H12LP Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures.

Codes / lots: Model Lot Manufacture Date Expiration H12LP 3 5 3 6 585 3/3/2015 Mar-18 H12LP 3 5 3 6 880 3/3/2015 Mar-18 H12LP 3 5 3 9 579 3/3/2015 Mar-18 H12LP 3 5 4 0 175 3/5/2015 Mar-18 H12LP 3 5 4 2 016 3/5/2015 Mar-18 H12LP 3 5 4 3 606 3/5/2015 Mar-18 H12LP 3 5 4 3 610 3/9/2015 Mar-18 H12LP 3 5 4 6 115 3/9/2015 Mar-18 H12LP 3 5 4 6 751 3/9/2015 Mar-18 H12LP 3 5 4 6 767 3/9/2015 Mar-18 H12LP 3 5 4 8 518 3/11/2015 Mar-18 H12LP 3 5 4 8 368 3/13/2015 Mar-18 H12LP 3 5 4 9 294 3/13/2015 Mar-18 H12LP 3 5 4 9 583 3/13/2015 Mar-18 H12LP 3 5 5 7 373 3/16/2015 Mar-18 H12LP 3 5 5 9 424 3/16/2015 Mar-18 H12LP 3 5 5 9 453 3/16/2015 Mar-18 H12LP 3 5 6 0 044 3/16/2015 Mar-18 H12LP 3 5 6 0 160 3/16/2015 Mar-18 H12LP 3 5 6 1 254 3/16/2015 Mar-18 H12LP 3 5 6 1 368 3/16/2015 Mar-18 H12LP 3 5 6 1 797 3/16/2015 Mar-18 H12LP 3 5 6 2 147 3/16/2015 Mar-18 H12LP 3 5 6 2 759 3/16/2015 Mar-18 H12LP 3 5 6 2 771 3/16/2015 Mar-18 H12LP 3 5 6 4 013 3/16/2015 Mar-18 H12LP 3 5 6 7 588 3/20/2015 Mar-18 H12LP 3 5 6 9 977 3/20/2015 Mar-18 H12LP 3 5 7 2 483 3/20/2015 Mar-18 H12LP 3 5 7 3 055 3/20/2015 Mar-18 H12LP 3 5 7 4 237 3/23/2015 Mar-18 H12LP 3 5 7 4 758 3/23/2015 Mar-18 H12LP 3 5 7 6 241 3/26/2015 Mar-18 H12LP 3 5 7 6…

Reason for recall

An increase in reports indicating that the reprocessed H12LP Trocar adjustable plug may break in the packaging or during use.

FDA-determined cause: No Marketing Application

Action taken

Stryker sent an Urgent Medical Device Recall letter to all affected customers on November 11, 2016. Customers were instructed to discontinue use of the affected lots and were asked to follow the link below to complete the Recall Effectiveness Check Form and indicate if any devices from the lots referenced in Attachment 1 remain in their inventory. This form must be completed even if no affected product is found. www.novasyte.com/sss142 Use the following reference number as you complete the form: [Ship to Account #]. If the form indicates that any affected devices remain in your inventory, a prepaid shipping label will be issued for the return of the product. Credit for all affected devices returned will be issued to your facility. Address any questions regarding the recall to Novasyte by calling (760) 279-3180 or emailing sss142@novasyte.com.

Details

Recall number
Z-0765-2017
Initiated
November 11, 2016
Posted by FDA
December 14, 2016
Terminated
May 5, 2017
Product code
NLM – Laparoscope, General & Plastic Surgery, Reprocessed
Quantity
40,476 devices
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Israel and Canada Foreign distribution to the following; IL and CA
Recalling firm
Stryker Sustainability Solutions, Tempe, AZ

Official FDA recall record