Recall Z-0775-2017
Angiodynamics, Inc. · initiated September 28, 2016
Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws
Product
ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515, Catalog No.REF 06597051, 2) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO S PG, Item Number H787065970535, Catalog No.REF 06597053, 3) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO A PG, Item Number H787065970555, Catalog No.REF 06597055 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
Codes / lots: Batch/Lot: 1) 5030537, 5037976, 5043444, 5059984, 5066413, 2) 5024545, 5031723, 5038659, 5042534, 5065538, 5071294, 5076428, 5082169, 3) 5022318, 5032456, 5034279, 5081986, 5082446
Reason for recall
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
FDA-determined cause: Component design/selection
Action taken
Angiiodynamics sent an Urgent Voluntary Medical Device Recall letter dated September 28, 2016, to all affected customers. Recall notifications were delivered by Federal Express. Customers were instructed to segregate and return all affected devices to AngioDynamics and to complete and return the Reply Verification Tracking Form, provided in the recall notification by fax to 1-800-782-1357. Customers were encouraged to forward a copy of the recall notification to all sites to which have distributed affected product. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form via to recall@angiodynamics.com. Customers with questions should contact customer service at 1-800-772-6446.
Details
- Recall number
- Z-0775-2017
- Initiated
- September 28, 2016
- Terminated
- June 18, 2018
- Product code
- DYB – Introducer, Catheter
- Quantity
- 996 boxes
- Distribution
- Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
- Recalling firm
- Angiodynamics, Inc., Queensbury, NY
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0772-2017 Class III | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |
| Z-0773-2017 Class III | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |
| Z-0774-2017 Class III | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |
| Z-0776-2017 Class III | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |
| Z-0777-2017 Class II | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |
| Z-0778-2017 Class III | ANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF…Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the… | 09/28/2016Terminated |